5.3.2. AUTOMATIC PRIMING AND SELF LEARNING PROCEDURE – NEW BORN CONFIGURATION
After manual priming is possible to continue with the automatic priming operation which also perform a centrifugal
pump self-learning procedure in order to improve the centrifugal pump safety and performances.
Warning
Automatic Priming must be executed once the circuit has been correctly filled with priming
solution.
Empty circuit might cause disposable centrifugal pump rupture or device malfunctions.
START / STOP PROCEDURE
You can select "START PRIMING" in order to start the
automatic procedure.
ECMOLIFE will set and hold for few minutes different
RPM targets within the range 0-4500RPM.
During this phase, the circuit will be automatically
primed and the centrifugal pump impeller will be
further verified in a self-learning procedure.
In case of emergency, procedure can be stopped and
restarted many time as necessary.
If the procedure is interrupted and restarted while in
progress, already verified set-point won't be executed
again.
Priming procedure requires a maximum of 4 minutes.
Once completed, it is possible to continue with the
patient connection.
SKIP PROCEDURE – ADULT/PAEDIATRIC/NEW BORN CONFIGURATION
In case of urgency you can SKIP the automatic procedure and complete the priming phase by manually setting the
RPM from the main screen.
Skipping the automatic procedure doesn't allow the device to perform the impeller self-learning operation.
For safety reason a RPM upper range limitation will be set for RPM values higher than 3500RPM and 4500RPM.
In order to reach an RPM set-point higher than 3500RPM and 4500RPM it's necessary to allow the device to perform
a two minutes of calibration in a range between these two target.
An alarm message will appear to inform the calibration is running and after a few minutes it will be possible to
increase the RPM setpoint. The alarm message and the yellow bar will disappear at the end of calibration
procedure.
When 3450 RPM or 4500 RPM setpoint is reached, the calibration will be performed by console by remaining for at
least 2 minutes in a range within 1550RPM-3450RPM for the first self-learning setpoint and 4300RPM-4450RPM
for the second self-learning setpoint.
Once both of the 3500RPM and 4500RPM setpoint has been validated, the device can be set in all the 0-5000 RPM
range for Adult/Paediatric configuration or 0-4500 RPM range for New Born configuration.
Warning
The treatment must be executed once the circuit has been correctly filled with priming solution.
Empty circuit might cause disposable centrifugal pump rupture or device malfunctions.