8. POWER SUPPLY

8.1. INTERNAL BATTERY

The device can work with both mains and internal battery powers.
In battery operation mode, in full charged condition, the guaranteed time for the device to function correctly is at
least 200 min.
The device automatically switches to battery operation when the power supply is interrupted. When the external
main power supply is available again, the device automatically returns to the external power supply. The internal
batteries are automatically charged when the device is connected to external main power supply.
When the device is stored or not used, it can be left connected to the external current to preserve the full charge
of the batteries.
The internal battery can be recharged while the sections are turned off.
Warning
If you decide to recharge the COLIBRÌ Console during the use of the device, internal battery may
require longer time to fully recharge depending on the treatment power consumption.

8.2. EXTERNAL BATTERY

During transport or in case main supply is not available, the device can be powered from the COLIBRÌ Auxiliary
Battery kit EU3968 from their relevant external battery sockets located on the COLIBRÌ right panel (1-2).
Once connected, the device will be powered from the external source, which will guarantee the charge retention of
the system for at least 200 minutes.