7.2. FUNCTIONALITIES
7.2.1. TURNING ON THE ECMOLIFE-COLIBRI' SYSTEM
In the Select Modality page of ECMOLIFE System select between "COLIBRÌ Configuration".
It is possible to connect the COLIBRI System to
ECMOLIFE in any treatment phase. Likewise, it is
possible to regain control on COLIBRI disconnecting
the connection cable or through the "COLIBRI" menu
on ECMOLIFE System at any time
At power on, ECMOLIFE System is able to detect if
the connection cable is connected to COLIBRI and if
COLIBRI is turned off or turned on. If colibrì is turned
off, ECMOLIFE will information about how to start
the treatment on COLIBRI System before switch to
ECMOLIFE control.
7.2.2. DIAGNOSTIC PHASE
If the ECMOLIFE-COLIBRÌ is correctly connected, in
ECMOLIFE SYSTEM diagnostic screen the device
will check colibri device is properly connected.
To start the diagnostic phase, press Start Autotest
button.
To retry the check, press Retry button.
During the diagnostic check the motor driver, the
pressure sensors and the flowmeter are related to
Colibrì system. Otherwise, venous probe and bubble
sensor are related to ECMOLIFE system.
During the Diagnostic phase the audio alarm of
ECMOLIFE system is checked and the acoustic
detection shall be confirmed manually.
If you want go back to previously page, press the
button displayed on the left of the screen.
Warning
-
Do not continue in case of acoustic alarm is not detected at the end of this phase. ALARM
SYSTEM might be compromised.
-
Before starting the diagnostic phase, check that the flowmeter is correctly connected to
the Colibrì System
Caution
When the treatment control switches from Colibrì to ECMOLIFE, the Colibrì device will entry in a
stand-by condition.
By pressing Start Treatment, all action will only be possible using the ECMOLIFE display
When the AUTOTEST is performed and the "Start Treatment" softkey is pressed, the ECMOLIFE System takes the
control of COLIBRI treatment.
When COLIBRI is connected to ECMOLIFE System, it is no longer possible to control the treatment through the
COLIBRI screen: treatment flow and RPM information will be displayed, together with colibrì charge level.
"Refer to Ecmolife" will be displayed on screen.
For other information related to treatment, alarms, parameters and more, refer to EU10583 IFU ECMOLIFE
SYSTEM.