BUBBLE SENSOR SCREEN
Select Bubble sensor menu you can
enable or disable the air bubble
detection.
By
default,
detection is not active.
The Menu Icon will be update in
accordance with the bubble detection
status:
ALARMS
In this section you can enter the Alarm Screen and configure RPM and LPM parameters alarm ranges
where you can establish alarms minimum and maximum threshold value. Out of the set limits the
device sends an alarm signal.
NAVIGATION BUTTON
Using the Navigation you can monitor the information related to power source.
The power source information are displayed in the toolbar.
The following information are available:
1. Integrated battery: charging/not charging and
battery level
2. External battery #1: connected/not connected and
battery level,
3. External battery #2: connected/not connected and
battery level,
4. 24V/ECMOLIFE: connected/not connected.
Warning
•
Check the generation of audible alarm signals. In case of not generated audio signal the use of
the device must be interrupted.
•
Suspend the use of the device should the audio alarm system does not function properly during
start-up test.
•
Setting audio alarm to maximum/minimum values could disable the proper detection of
physiological alarms.