3. FLOWMETER / AIR SENSOR CONNECTION ON THE TUBING
Caution
Check that the sensor is free from contamination or dust.
The Flowmeter/air bubble sensor must be located on the patient return line at a distance of at least 50 cm from
inlet of the cannula.
Warning
•
A backup flowmeter must be available in proximity of the device, next to the console.
3.1. COLIBRI AND LANDING MONITOR FLOW MEASUREMENT
In order to achieve an adequate correspondence between the LANDING monitor flow measure and COLIBRÌ console
flow measure make sure that the two flowmeters are attached to a 15cm straight tube before and after the
transducers.
Check the indications on the EUROSETS tubing set to identify the suggested flowmeter positioning, in order to
minimize the flow turbulence and the associated measurement inaccuracies related to the sensor.