PARAMETERS EDITING
COLIBRÌ Console provides 4 membrane buttons:
1. Increase parameter value;
2. Decrease parameter value;
3. Return button;
4. Lock Screen button.
Parameter selection
You can activate the parameter selection that you
want change, pushing for at least 1 second directly on
the value that you want to modify on the touchscreen.
Increase and decrease values
Using the +/- selection display buttons it is possible to increase / decrease the value of the selected parameter.
Edit confirmation or refusal
Following the treatment parameter editing (e.g. RPM, LPM) it is necessary to confirm the set value by re-pushing
the selected value on the display, within at least 3 seconds.
The parameter editing will be performed only if the user confirms.
If you delete the operation using the RETURN button, or if it takes more than 3 seconds to confirm, the device will
restore the value to the previous one.
SCREEN LOCK
Automatic lock screen
Automatic screen lock: to avoid editing of unwanted settings or functions, all display buttons will be automatically
locked after 5 minutes of inactivity.
You can reactivate the touchscreen by pushing the unlock button.
The touchscreen will be reactivated in case of alarm. If the user does not intervene within the set time limit, the
touch will be deactivated.