System
Show system info. HW and SW version.
Download
In this section you can download the patient data and save it into any memory drive by connecting
an USB key to the Primary USB key connection in the front panel. During the treatment, is possible
to download the last treatment performed and/or the treatment ongoing.
Before starting the treatment, it is possible to download the last treatment log file or the complete
log file, with all the treatment data performed until then.
Download process duration for the complete log file might takes several minutes, depending on the
amount of data stored inside the console memory.
Data can be downloaded both at the end of the treatment and during the treatment (sampled data
will be saved up to that point).
The user will receive a confirmation of data successful transfer.
To access to log data, please contact the manufacturer authorized technical assistance service.
PRESSURE AND BUBBLE SENSOR MENU
In this section you can enter the
Pressure sensor and Bubble sensor
menu.
PRESSURE SCREEN
Selecting Pressure sensor menu you
can monitor the following information:
In this screen, you can also proceed
with the zero pressure procedure.
Warning
Calibrate external and integrated pressure sensors before use.
en
Blood Flow (LPM),
Pump Speed (RPM),
Drainage pressure (Pdrain),
Pre-oxygenator
pressure
(Pin),
Post-oxygenator
pressure
(Pout),
Pressure drop (DP)
21