13. MAINTENANCE - SERVICING - SYSTEM LIFESPAN

13.1.

LIFECYCLE OF THE RECHARGEABLE INTERNAL BATTERY

The battery pack must be changed every 36 months at the time of the third scheduled maintenance.
13.2.

PERIODICAL MAINTENANCE PROGRAM

COLIBRÌ System shall undergo periodical maintenance at 12 months intervals.
The maintenance shall be carried out exclusively by manufacturer authorized personnel.
Warning
No servicing action is admitted on COLIBRÌ System when the device is operating or when the
patient is connected to it.
Check the integrity of visible insulation and do not use the device if the equipment is damaged.
It is responsibility of the healthcare facility using the device to ask for technical service to the manufacturer or
authorized agent.
13.3.

POSSIBLE ANOMALIES OCCURRING DURING DEVICE FUNCTIONING

Whenever a device malfunctioning occurs, suspected or communicated through display messages, the user shall
contact the manufacturer or manufacturer authorized representative.
The operator is not authorized to carry out servicing or calibration operations on the device.
It is allowed to perform adjustments of venous probes and pressure sensors as indicated in the present user
manual.
The operator is responsible for checking device integrity before and during each use.
Warning
Whenever the operator suspects a device component failure, device use shall be immediately
discontinued and manufacturer authorized technical assistance service contacted.
Check the integrity of visible insulation and do not use the device if damaged.
13.4.

MEDICAL EQUIPMENT LIFESPAN

Assumed that the maintenance program is observed, the medical equipment has a lifespan of 10 years from the
delivery date to the final customer / healthcare facility.
Warning
Essential performance and safety are not guaranteed in case of:
Maintenance actions (included battery pack replacement) and hardware modifications not
performed by manufacturer or authorized representative.
Missed observation of the required periodical maintenance program by the manufacturer or
authorized representative.