5.2. TOOLBAR
Parameters displayed in the main screen are listed below:
Parameter
Description
Detected blood
Blood flow
flow speed
Blood flow
Detected blood
(NEW
flow speed
BORN)
Pump speed
Speed
revolutions per
pump
minut
Speed
Pump speed
pump
revolutions per
(NEW
minut
BORN)
P
Suction pressure
drain
Inlet pressure in
oxygenator module
P
and outlet
in
pressure from
driver engine
Outlet pressure
P
from oxygenator
out
module
en
Measure
Sensor
range
Colibrì
FLOWMETER 3/8"
0,0 ÷ 10,0
x 3/32""
Colibrì
FLOWMETER
0,0 ÷ 3,0
1/4''x3/32''
/
0 ÷ 5000
/
0 ÷ 4500
Single use pressure
-200 ÷
sensor
+600
Single use pressure
-200 ÷
sensor
+600
Single use pressure
-200 ÷
sensor
+600
Measure unit
Accuracy
±0.07 l/min for flow
value <= 1.0 l/min
LPM
± 7% for flow value
> 1.0 l/min
±0.07 l/min for flow
value <= 1.0 l/min
LPM
± 7% for flow value
> 1.0 l/min
RPM
± 50 RPM
RPM
± 50 RPM
± 10mmHg for
pressure value < -50
mmHg
± 5mmHg for -50
mmHg <= pressure
mmHg
value <= +50 mmHg
(convertible in
± 10mmHg for
kPa)
+100 mmHg <
pressure value <=
+300 mmHg
± 15mmHg for
pressure value >
+300 mmHg
± 10mmHg for
pressure value < -50
mmHg
± 5mmHg for -50
mmHg <= pressure
mmHg
value <= +50 mmHg
(convertible in
± 10mmHg for
kPa)
+100 mmHg <
pressure value <=
+300 mmHg
± 15mmHg for
pressure value >
+300 mmHg
± 10mmHg for
pressure value < -50
mmHg
± 5mmHg for -50
mmHg <= pressure
mmHg
value <= +50 mmHg
(convertible in
± 10mmHg for
kPa)
+100 mmHg <
pressure value <=
+300 mmHg
± 15mmHg for
pressure value >
+300 mmHg
23