11. TECHNICAL SPECIFICATIONS
Regulations
Classification
Reference standards
Conditions of use
Conditions of use
Safety of use in the presence of
flammable anesthetic mixtures
with air or oxygen or nitrous
oxide:
Environment
46
Classified in accordance with Regulation (EU) 2017/745 (MDR)
•
IEC 60601-1:2005+AMD1:2012+AMD2:2020
•
IEC 60601-1-2: 2014+AMD1:2020
•
IEC 60601-1-8:2006+AMD1:2012+AMD2:2020
•
IEC 62304:2006 + AMD1:2015
•
IEC 60601-1-6:2010+AMD1:2013+AMD2:2020
•
IEC 60601-1-12:2014+AMD1:2020
•
IEC 62366-1:2015+AMD1:2020
•
IEC 62353:2014
•
EN ISO 14971:2019
•
UNI EN 1789:2020
•
BS EN IEC 62311:2020
•
ETSI EN 301 489-1 V2.2.3 2019
•
ETSI EN 301 489-17 V3.2.4 2020
•
ETSI EN 300 328 V2.2.2 2019
•
UNI EN13718-1 2014 + A1 2020
•
RTCA DO160G: section 4, 6, 7, 8, 20, 21, 26 - only AW139 helicopter model
Device for continuous operation
Device not suitable for use in the presence of flammable anesthetic mixtures
with air or oxygen or nitrous oxide.
During operation:
•
Temperature:10 to +40°C,
•
Humidity: 15-90%,
•
Pressure range of 760-1060 hPa.
During transient operation:
•
Temperature:10 to +40°C,
•
Humidity: 15-90%,
•
Pressure range of 760-1060 hPa.
During storage/transport, when the device is not operating:
•
Temperature:-18 to +50°C,
•
Humidity: 15-90%,
•
Pressure range of 620-1060 hPa.