Download Print this page

Eurosets COLIBRI SYSTEM Instructions For Use Manual page 13

Advertisement

COLIBRÌ System primary unit has integrated electronic to measure the following parameters:
Patient drainage pressure (Pdrain),
Blood flow (LPM),
Pump speed (RPM),
Pre-oxygenator pressure (Pin),
Post-oxygenator pressure (Pout),
Air bubble presence in the blood return tube by means of Flowmeter.
Warning
Make sure that the power supply is available and the battery charge is full.
The protective cap must always be placed in the relevant connectors if they are not used.
Do not allow tubes or cables to hang down.
Ensure that there is no strain on tubes or cables.
Avoid kinking of tubes or cables.
Warning
An Eurosets centrifugal pump backup unit (as COLIBRÌ System and/or ECMOLIFE System)
must be available in proximity of the device, next to the console.
In order to make sure the backup device works correctly in case of emergency, switch it on
and perform the diagnostic phase at least once every 48h while using the primary device.
Check the activation of Alarm LED. In case of failure in the activation the use of the device
must be interrupted.
Eurosets disclaims the relevant risk of a use of a device with backup Colibrì not checked.
Caution
Place the equipment in order to not cover the air intakes to allow a proper heat dissipation.
en
13

Advertisement

loading
Need help?

Need help?

Do you have a question about the COLIBRI SYSTEM and is the answer not in the manual?

Related Products for Eurosets COLIBRI SYSTEM