6. ALARMS

6.1. ALARM RANGE AND PARAMETERS

It is possible to set the alarm limits in the Alarm screen.
The alarm activates whenever the parameter value exceeds the warning limits, i.e:
Value measured above the upper limit
Value measured above the lower limit
You can set the alarm limits within the parameters summarized in the following table.
6.1.1. ALARM LIMIT – ADULT/PAEDIATRIC CONFIGURATION
Parameter
Description
Detected
Blood flow
blood flow
speed
Pump speed
Speed
revolutions
pump
per minute
Suction
P
drain
pressure
Inlet
pressure in
oxygenator
module and
P
in
outlet
pressure
from driver
engine
Outlet
pressure
P
from
out
oxygenator
module
Air bubble
Bubble
detection
en
Measure
Sensor
range
Clamp on
transducer
0,0 ÷
Sct 3/8" x
10,0
3/32"
/
0 ÷ 5000
Single use
-200 ÷
pressure
+600
sensor
Single use
-200 ÷
pressure
+600
sensor
Single use
-200 ÷
pressure
+600
sensor
Bubble
> 65
sensor
Measure
Alarm
unit
threshold
LPM
0,0 ÷ 10,0
RPM
0 ÷ 5000
mmHg
-200 ÷
(convertible
+600
in kPa)
mmHg
-200 ÷
(convertible
+600
in kPa)
mmHg
-200 ÷
(convertible
+600
in kPa)
(not
µl
modifiable)
Default
Preliminary
alarm
Verification (*)
threshold
Check no
occlusions or
any other case
0,2 ÷ 8,0
which may
compromise the
flow are
presents.
Check no
occlusions or
any other case
2000 ÷
which may
4500
compromise the
flow are
presents.
Check for and
resolve a
physiological or
mechanical
-80 ÷ 25
cause. Ensure
appropriate
pressure alarms
are set.
Check for and
resolve a
physiological or
mechanical
0 ÷ 500
cause. Ensure
appropriate
pressure alarms
are set.
Check for and
resolve a
physiological or
mechanical
0 ÷ 500
cause. Ensure
appropriate
pressure alarms
are set.
Check any
leakage on
> 65
tubing set are
present.
29