12. ELECTROMAGNETIC COMPATIBILITY
COLIBRÌ System is intended for use in the electromagnetic environment specified below. User shall ensure that
COLIBRÌ System is used in such environment.
Guidance and manufacturer's declaration – electromagnetic emissions
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions EN 61000-3-2
Voltage fluctuations/flicker
emissions EN 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
Immunity Test
Electrostatic discharge
(ESD) EN 61000-4-2
Electrical fast
transient/burst
EN 61000-4-4
Surge EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines EN
61000-4-11
Power frequency (50/60
Hz) magnetic field
EN 61000-4-8
Radiated fields in close
proximity EN 61000-4-39
Radiated, radio-
frequency,
electromagnetic field
EN 61000-4-3
(*) Un is the a.c. mains voltage prior to application of the test level.
en
Compliance
Group 1
Class A
Class A
Complies
EN 60601-1-2 Test level
±8 kV contact
±2, ±4, ±8, ±15 kV air
±2 kV for power supply lines
100 kHz repetition rate
±1 kV differential mode
±2 kV common mode
0% Un (*) for 0.5 cycles
70% Un (*) for 25 cycles
0% Un (*) for 250 cycles
30 A/m
8 A/m for 30 kHz
65 A/m for 134.2 kHz
7.5 A/m for 13.56 MHz
3 V/m for 80 MHz – 1 GHz
27 V/m for 380 MHz – 390 MHz
28 V/m for 430 MHz – 470 MHz
9 V/m for 740 MHz – 787 MHz
28 V/m for 800 MHz – 960 MHz
3 V/m for 1 GHz – 2.7 GHz
28 V/m for 1.7 GHz – 1.99 GHz
28 V/m for 2.4 GHz – 2.57 GHz
9 V/m for 5.1 GHz – 5.8 GHz
Electromagnetic Environment
COLIBRÌ System uses RF energy
only for its internal function.
Therefore, the RF emission is very
low and not likely to cause any
interference in nearby electronic
equipment
COLIBRÌ System is suitable for use
in all establishments other than
domestic and those directly
connected to the low voltage
power supply network which
supplies buildings used for
domestic purposes
Compliance level
EN 60601-1-2 Test
level
EN 60601-1-2 Test
level
EN 60601-1-2 Test
level
EN 60601-1-2 Test
level
EN 60601-1-2 Test
level
EN 60601-1-2 Test
level
EN 60601-1-2 Test
level
Electromagnetic
Environment
Hospital
Hospital
Hospital
Hospital
Hospital
Hospital
Hospital
49