5    Maintenance
NOTICE!
Improper mounting or maintenance of Sip-N-Puff control 
may cause damage to input module by water or saliva.
— Mouthpiece and breath tube MUST be completely dry 
before installation.
5.1    Replacing Mouthpiece
 1.  Remove mouthpiece A from gooseneck 
B.
A
Make sure to leave lipswitch C in shrink 
B
C
sleeving which keeps together lipswitch 
and mouthpiece.                        
 2.  Insert new mouthpiece.                        
5.2    Replacing Saliva Trap
NOTICE!
If saliva trap is inserted wrong way round, input module can 
get damaged by water or saliva.
— Make sure to insert saliva trap in correct orientation.
— Saliva trap MUST be installed to reduce risk of water or 
saliva entering input module.
1637423-K
1.
B
A
2.
C
3. Insert new saliva trap with INLET imprinting facing towards input 
module.
5.3    Cleaning Sip-N-Puff
Cleaning at least twice a week is recommended.
A
B
C
 2.  Remove breath tube from saliva trap, refer to 5.2 Replacing
Saliva Trap, page 91.              
 3.  Position catch can beneath breath tube to collect water                 a nd 
rinse.              
 4.  Flush mouthpiece and breath tube with warm running water.              
 5.  Rinse with oral rinse to disinfect.              
 6.  Let dry completely before installation.              
 7.  Install mouthpiece, lipswitch and breath tube.              
5    Maintenance
Remove screw / hand screw A and 
backrest shroud B.
Remove saliva trap C from tube.
 1.  Remove mouthpiece A and 
lipswitch C from gooseneck B.          
91