Changing Function Cards
The lipswitch can also be used as a mode function switch.
For difference between function card and profile, refer to 2.2.3 User Function Card Overview, page 8.
 1.  Stop wheelchair.                                
 2.  Short press lipswitch to change function card.                                
 3.  Long press lipswitch to change profile.                                
4.21.7    Using the Head Array
WARNING!
Risk of serious injury
Proximity sensors are sensitive to water. If enough water is present close to sensors, they may be activated and power wheelchair may 
start moving unintentionally.
— Do not operate Head Array with wet hair.
— Do not operate Head Array in wet weather.
— Do not operate Head Array in any circumstances where water may come close to sensors.
WARNING!
Risk of serious injury
Sensor pads are made of water resistant vinyl to get water quickly run off the pads before activating sensors.
If sensor pads are damaged, water may get in and power wheelchair may start driving unintentionally.
If sensor pads are covered by water absorbing material, power wheelchair may start driving unintentionally.
— Do not operate Head Array if sensor pads are damaged. Change sensor pads immediately.
— Do not cover sensor pads with any material.
1637423-K
Seating functions can only be operated with the right or left             p ad 
of the head array.
4    Usage
81