4
Basic Operation

Starting Up and Shutting Down the Instrument

Starting Up and Shutting Down the Instrument

Starting in Standby mode

1 Check that the nitrogen gas for the drying gas and for the collision cell are
2 Turn on the power for the computer and monitor.
3 Check that the instrument and rough pump are plugged into a power outlet.
4 Press the front power switch on the instrument.

Turning off the instrument

Before you turn off the instrument, complete any acquisition steps and save your
data.
To prevent instrument damage, always vent the system before pressing the
C AU T I O N
power button. Only the Lab Manager and Everything role has access to venting.
Do not touch the spray shield and related spray chamber components. They
WAR N IN G
are likely to be hot.
1 Turn off the solvent flow.
2 In Acquisition, in the Method Editor window, click the MS Q-TOF tab.
3 Click Tune > Advanced Tune in the left pane.
4 Click
5 Click
Agilent Revident LC/Q-TOF System User Guide
turned on.
The vacuum system automatically starts to pump down the instrument, and
the electronics are turned on. The system card boots up and the firmware
starts.
If there is any analyte in any of the components in the sample delivery system,
flush the delivery system with pure solvent before continuing.
Tune control.
Launch Advanced Tune.