Maintenance

Disinfection

Disinfection

Schedule
To prevent the spread of respiratory illness, disinfect your instrument as needed.
Equipment List
Steps
1 Put on personal protective gear (gloves, mask, gown).
2 Make sure that your instrument is cooled to room temperature.
3 Dampen a clean lint-free cloth with the 70:30 isopropyl alcohol:water mix.
4 Gently wipe external surfaces to be cleaned using moistened cloth. Do not
5 Wipe the outer surface of cables and their connectors but do not touch the
6 Use one cloth per instrument to prevent cross contamination.
7 Dispose of the cloth appropriately.
8 Allow all surfaces to completely air-dry. No moisture should be present on the
9 Discard the gloves and other personal protective equipment or clean them in
Agilent Revident LC/Q-TOF System User Guide
• Clean, powder-free nitrile gloves
• Clean lint-free cloth
• Gown
• Isopropanol
• Mask
allow any liquid to drip into the instrument.
electrical connections (for example pins and plugs inside the cable or
connection port).
instrument surfaces.
an approved process, then wash your hands.