Maintenance

Inspecting the nebulizer

Inspecting the nebulizer

Schedule
Perform this procedure if you have the G1960-67470 Nebulizer Adjustment Kit.
Perform this project if there is increased LC backpressure, off-axis spraying,
dripping from the nebulizer, or when the reference nebulizer needle is not
spraying.
Equipment List
Parts needed:
Steps
1 Determine your nebulizer type per the user guide or the document that comes
2 Carefully insert the tip of the nebulizer into the end of the Nebulizer
3 Slide the heads of the two spring loaded socket head cap screws through the
4 Turn the nebulizer body 1/8 turn counterclockwise.
If the nebulizer is too tight or too loose, loosen or tighten the nut on the captured
NOTE
screws to adjust the tension on the spring. Turn clock wise to tighten, counter
clock wise to loosen.
5 Gently push the 40 x Magnifier into the end horizontal position of the
6 Push the LED Flashlight into the Nebulizer Adjustment Fixture. No tool is
7 Turn on the LED Flashlight. The on/off switch is at the end of the flashlight.
98
• Clean powder-free nitrile gloves
• G1960-67470 Nebulizer Adjustment Kit
• T6 Torx driver
• 1/4-inch x 5/16-inch open-end wrench (Quantity: 2)
• Replacement needed kit
with the kit (need the part number).
Adjustment Fixture.
mounting holes on the body of the nebulizer.
Nebulizer Adjustment Fixture. Align the tapered tip of the magnifier to the tip
of the nebulizer needle.
needed.
Agilent Revident LC/Q-TOF System User Guide