Select the Q-TOF/TOF ion source

Select the Q-TOF/TOF ion source

The Q-TOF/TOF ion source is automatically detected and displayed in the
System Settings pane of the Acquisition Section of the Method Editor and in the
Q-TOF dashboard tile in the Instrument Status pane.
If you are using the Offline Method Editor, select the desired ion source from the
Method Editor window. To open the Offline Method Editor, select an LC/MS
instrument in Agilent Control Panel, and click Launch Offline, or double-click in
the leftmost column of a row in the Worklist window. An acquisition method
must be specified to open the Offline Method Editor from this window.
1 To display the Method Editor window, click Method Editor in the windows
2 Select the MS Q-TOF tab.
3 Select the Method > Source section.
4 From the Ion Source drop-down list, select the ion source.
5 Click Apply Method in the Method Editor toolbar.
Agilent Revident LC/Q-TOF System User Guide
section on the Ribbon. Or, click the Method layout in the Ribbon.