Hardware Overview

Instrument Status Indicators in Acquisition

Instrument Status Indicators in Acquisition

In the Acquisition program, the Instrument Status window indicates the state of
the instrument.
Gray
Yellow - Unknown: The device is in an unknown run state.
Light Orange - Not Ready: The instrument is connected but is not ready to run
(due to not reaching the correct temperature or pressure required by the method,
for example).
Green - Idle: The instrument is on and ready to process samples.
Teal - Standby: The instrument is in a standby/sleep state. Put the instrument
in Standby mode when it is not in use or to change the ion source. When the
instrument is in Standby mode:
Magenta - Prerun/Injecting: The instrument is on and is preparing to start
acquisition.
Blue - Running, postrun: The instrument is currently collecting data.
Orange - Tuning: The instrument is in the process of adjusting MS parameters.
Red - Instrument Error: The instrument has an error and cannot process
samples.
Agilent Revident LC/Q-TOF System User Guide
• Offline: The instrument is configured with the system and available to use,
but the AIC/workstation is not currently running. The amount of time for an
instrument to reach the Idle state depends on each instrument.
• Disconnected: The connection to the instrument has been closed.
The source interface is on, the drying gas remains heated, and nebulizer
flows are maintained.
The source and ion optics voltages are turned off, and the mass
spectrometer is not sending spectra to the computer.
For APCI and multimode sources: Vaporizer gas heaters are turned down.
For the Agilent Jet Stream (AJS) source: The sheath gas temperature,
drying gas flow, and drying gas temperature are reduced.
The spray chamber high voltages are turned off.
The mass spectrometer stops generating spectra.
23