Tuning

Mass Calibration

Mass Calibration

Mass Calibration performs a TOF mass calibration on the Active Instrument
Mode, which is displayed in the Systems Settings panel. It assigns masses based
on known masses of standard compounds. Agilent recommends running mass
calibration daily or weekly, at a minimum for optimal precision and accuracy.
During normal operation, mass calibration lasts approximately 5 minutes.
Only the active instrument mode can be calibrated.
NOTE
1 In the Method Editor window, click the MS Q-TOF tab.
2 Click Tune > Mass Calibration in the left pane.
3 Click
4 Select Mass Calibration.
5 Click
6 Click
Agilent Revident LC/Q-TOF System User Guide
Tune control.
Mass calibration/Checktune. When the tune completes, review the
report.
Tune control to release control of the LC/Q-TOF instrument.