Internal Reference Mass Correction
For those sources without a reference sprayer (APCI), there are two options: (1)
NOTE
Use the UIRM-hardware kit (p/n G1969-67400.) (2) For other sources that do not
have a second sprayer, connect a tee into the LC tubing just before the sprayer
and admit a reference mass solution into the tee using a syringe pump or other
pumping device.
4 Enter the desired Reference parameters for Flowrate, Nebulizer (psi), and
5 In the Reference Masses table, select up to two reference standard masses
Use default values or choose two masses spread across the application mass
NOTE
range. To add new values to the Reference Masses table, click
end and enter the exact mass values.
6 When finished, save the acquisition method.
Agilent Revident LC/Q-TOF System User Guide
Reference Mass.
for automatic mass correction.