Figure V-15

5.4.8 Settings reset

Settings Reset will return the DVNext settings (Device, User and Global) to the factory default. Internal memory will
not be affected. User accounts will not be affected.

5.4.9 Device reset

DATA LOSS
Device Reset will return the DVNext to the factory default condition. All data, tests and Audit Trails will be deleted
from internal memory. All User accounts will be deleted.
TIP: Move data files and test files to a USB Flash Drive prior to selecting Device Reset.

5.4.10 Calibration Reminder

The DVNext is programmed to provide the user with a reminder that calibration is due. Two parameters must be
set: Frequency in Months and Start Date. When both parameters are entered, the Calibration Reminder screen (see
Figure V-16) will calculate and display the day when the reminder will be presented.
Check the Calibration Reminder On check box to activate this feature.
The Calibration Reminder will be presented during the first power up on or after the specified day. The message box
(see Figure V-17) requires that the user press OK before the leveling/AutoZero will be performed.
The calibration reminder might be reset by AMETEK Brookfield or your local authorized dealer when the unit is
returned for calibration.
M19-2101 REVISION 2.0
D V N E X T R H E O M E T E R - O P E R AT I O N A L M A N U A L