APPENDIX N - FAULT DIAGNOSIS AND TROUBLESHOOTING

Listed are some of the more common problems that you may encounter while using your Rheometer.
Spindle does not rotate
Make sure the rheometer is plugged in.
Check the voltage rating on your rheometer (5, 220V); it must match the wall voltage.
Make sure the motor is ON and the desired RPM is selected.
Make sure the spindle is tightened securely to the rheometer coupling.
Check the straightness of all other spindles; replace if bent.
Inspect rheometer/spindle threaded coupling and mating areas for dirt; carefully clean threads in the
spindle coupling with a 3/56 tap.
Inspect rheometer/spindle magnetic couplings and mating areas for debris; e.g. metallic chips or shavings.
Inspect threads for wear; if the threads are worn, the unit needs service (see Appendix O). Check to see if
spindles rotate eccentrically or wobble. There is an allowable run-out of 1/32-inch in each direction (1/16-
inch total) when measured from the bottom of the spindle rotating in air.
Check to see if the rheometer coupling appears bent; if so, the unit is in need of service (see Appendix O:
Warranty Repair and Service).
Verify spindle, speed, and model selection.
Verify spindle selection is correct on DVNext.
If % readings are under-range (less than 10%), the display will flash; change spindle and/or speed.
"EEEE" on the digital display means the unit is over-range (greater than 100%); reduce speed and/or
change spindle.
Verify test parameters: temperature, container, volume, method. Refer to: "More Solutions to Sticky
Problems", Section III "DVNext Rheometer Operating Instructions", Appendix C: Variables in Viscosity
Measurements.
Perform a calibration check; follow the instructions in Appendix F.
Verify tolerances are calculated correctly.
Verify the calibration check procedures were followed exactly.
If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix O for details
on Warranty Repair and Service.
Rheometer will not return to zero
Rheometer is not level
Check with spindle out of the sample.
Adjust the laboratory stand.
Pivot point or jewel bearing faulty
Perform an Oscillation Check*
Refer to Appendix H for Automatic Oscillation Check.
*This procedure does not apply to instruments with ball bearing suspension (see Section I.3).
Display reading will not stabilize
Special characteristic of sample fluid. There is no problem with the rheometer.
Refer to Appendix C.
Check for erratic spindle rotation.