A.4 Verifying calibration
Determine the appropriate sample volume. Refer to Table A-1 to determine the correct sample volume
required for the spindle to be utilized.
Select a AMETEK Brookfield Viscosity Standard fluid that will give viscosity readings between 10% and
100% of full-scale range. Refer to Appendix B for viscosity ranges of cone spindles. AMETEK Brookfield
uses mineral oil viscosity standard fluids to calibrate Wells Brookfield Cone/Plate Rheometers at the
factory. AMETEK Brookfield recommends that customers use mineral oil viscosity standard fluids when
you perform a calibration check.
If you decide to use a silicone viscosity standard fluid, do not use a fluid with a viscosity value greater
than 5000 cP with a Cone/Plate. AMETEK Brookfield offers a complete range of mineral oil viscosity
standards suitable for use with Cone/Plates for viscosities above 5,000 cP or shear rates above 500
sec-1; see Table E-2 in Appendix F for a list of available fluids.
It is best to use a viscosity standard fluid that will be close to the maximum viscosity for a given cone
spindle/speed combination.
Example: DVNext LV Rheometer, Cone Spindle CPA-42Z, AMETEK Brookfield Viscosity Standard
having a viscosity of 9.7 cP at 25°C.
At 60 RPM, the full-scale viscosity range is 10.0 cP. Thus, the Rheometer reading should be 97%
torque and 9.7 cP viscosity ± 0.197 cP. The allowable error (±0.197 cP) is a combination of Rheometer
accuracy and fluid tolerance (refer to Interpretation of Calibration Test Results in Appendix F).
With the motor off, remove the sample cup and place the viscosity standard fluid into the cup.
Attach the sample cup to the Rheometer and allow enough time for the sample, cup and cone to reach
temperature equilibrium (25°C ±0.1°C).
Go to Configure Viscosity. Set the desired parameter, e.g. Speed, Spindle, Sample ID, etc. Set the End
Condition for Time 00:00:00. Press Run.
Measure the viscosity and record the readings in both % torque and centipoise (cP).
Verify that the viscosity reading is within the allowable 1% deviation, as explained earlier, for the specific
viscosity standard fluid(s) that you are using.
The CPE, CPM, or CPA designation on the cone spindle indicates use with Electronic Gap Setting Cone/
Plate Rheometers/Rheometers only.
M19-2101 REVISION 2.0
CONE PART NO.
CPA-40Z or CPM-40Z
CPA-41Z or CPM-41Z
CPA-42Z or CPM-42Z
CPA-51Z or CPM-51Z
CPA-52Z or CPM-52Z
D V N E X T R H E O M E T E R - O P E R AT I O N A L M A N U A L
Table
A-1
SAMPLE VOLUME
0.5 mL
2.0 mL
1.0 mL
0.5 mL
0.5 mL