Fix the pump to the rack with bolts at four places.
Fix the pump horizontally so that there is not a wobble.
Refer to "3.1 Storage/ Installation", too.
Model: PMB100D
① The connection of the pipe to the PMB 100Dserves as the installation of the pump.
The pipe should be able to support the weight of the pump(example: the pipe is welded to
the rack).
② The pump has four tapped holes at the outlet.
Tapped hole:M8, 14mm
The pump is fixed with four bolts.
Depending on the thickness of the flange on the pipe, please regulate the length of the bolts.
Length of engagement: More than 12mm.
xvii
YK11-0016-DI-002-08