System Shutdown, Startup, and Reset
Shutting Down the System in an Emergency
Thermo Scientific
When you are not using the TSQ Endura or TSQ Quantiva system for short periods of time,
place the mass spectrometer in standby mode. For longer periods, for example, two or more
months, you can shut it down completely. In addition, many maintenance procedures for the
system require shutting down the mass spectrometer completely.
Contents
•
Shutting Down the System in an Emergency
•
Placing the Mass Spectrometer in Standby Mode
•
Turning On the Mass Spectrometer
•
Shutting Down the Mass Spectrometer Completely
•
Starting the System After a Complete Shutdown
•
Resetting the Mass Spectrometer
•
Resetting Calibration Parameters
•
Restarting the Data System
•
On/Off Status for MS Components Under Varying Power Conditions
CAUTION If you must turn off the mass spectrometer in an emergency, turn off the main
power switch located on the right-side power panel
power to the mass spectrometer, including the forepump, without harming components
within the instrument. However, do not use this method as part of the standard shutdown
procedure. Instead, see
To turn off the LC, autosampler, and data system computer in an emergency, use their
respective on/off switch or button.
"Shutting Down the Mass Spectrometer Completely."
TSQ Endura and TSQ Quantiva Hardware Manual
(Figure
3). This switch turns off all