Changing The Hp-Ib Address; Choosing Hp 3325A Compatibility - HP 3324A Operating And Programming Manual

Synthesized function/sweep generator
Hide thumbs Also See for 3324A:
Table of Contents

Advertisement

Changing the
HP-B Address
Choosing HP
3325A
Compatibility
8-4 Utilities
If you wish to change the HP-IB address, scroll the
utilities menu until "HPIB addr.: xx" is shown. Type
in the address number required (0 to 30) and then press
(Enter).
The HP 3324A refuses to accept a new HP-IB address
if it is in the remote state, or if it is addressed either as
a listener or as a talker. Un-address the HP 3324A (via
the controller) before changing the HP-IB address.
The default HP-111 address is 17.
The HP 3324A can be set so that at power-on, it powers
up just like the HP 3325A. This means that in the
compatible mode the main output signal is enabled and
the instrument is reset to its default setting, including
clearing all of the instrument states stored. In the
non-compatible mode at power-on, the main output
signal is disabled, the instrument is turned on with the
settings used previous to power-down, and none of the
stores are cleared. Also, for the compatible mode log
intervals in a continuous sweep are always approximated
using two linear segments. In the non-compatible
mode for a continuous sweep, the log intervals are
approximated as in a single sweep for sweep times 21
s. For sweep times <l s (down to 0.1 s) the two linear
segment approximation is used.
To change the mode, display "compatib.: yes/NO" from
mode and press (Selectl.
shown in the uppercase and the other in the lowercase.
The activated mode is then