XMF; Multi-Interval
Marker Frequency
The XMF command sets the marker frequency for a
particular sweep interval. IXMF causes the instrument
to output the current marker frequency for a particular
sweep interval. The XMF and IXMF commands are
identical to the MF and IMF commands, respectively,
when an index of 1 is used.
Instrument Preset value: 5.0 MHz
Both commands are not available on the HP 3325A.
Syntax
1
to any of "SNI" Sets frequency of interval.
Table E-49. XMF "value" Range Given "units"
"value" Range (sine) "units" Description
0.0 to 21000000.000
0.0 to 21000.000000
0.0 to 21.000000
Table E-50. IXMF Response Format
Query Command
IXMF
v a l u e
u n i t s
Meaning
HZ
Hertz
Kilohertz
KH
MH
Megahertz
Response Format
HP 3324A Command Syntax E-45
.
.