Triangle Linearity; Specifications; Equipment Required; Procedure - HP 3324A Operating And Programming Manual

Synthesized function/sweep generator
Hide thumbs Also See for 3324A:
Table of Contents

Advertisement

Triangle Linearity

Specifications

Equipment Required

Procedure

12-34 Performance Tests
This procedure tests the linearity of the HP 3324A
triangle wave output. As the triangle and ramp outputs
are generated by the same circuits, this procedure
effectively tests the ramp linearity also.
High-speed DC Digital Voltmeter
Resistor, 2Ofi l/4 W 1%
Resistor, 30R l/4 W 1%
BNC-to-Triax
1. Connect the HP 3324A and the high-speed voltmeter
through the divider as shown in Figure 12-9.
Figure 12-9. Triangle Linearity Test Set-Up
Adapter
External
Trigger
DVM
20 Ohms