DC Offset
Accuracy (DC
Specifications
Equipment Required
Procedure 1
12-30 Performance Tests
This procedure tests the HP 3324A DC offset accuracy
when no AC function output is present.
1% of fullrange ho.02 mV
DC Digital Voltmeter
Feedthru Termination
Connect the HP 3324A signal output directly to the
the DC digital voltmeter input.
3. Set the HP 3324A DC offset to 5 V, then press
4. The voltmeter reading should be between i-4.950 and
5. Change the HP 3324A DC offset to -5 V. The
voltmeter reading should be between -4.950 and
- 5 . 0 5 0 v .