The HP 70340A/41A Quick Start Guide
To Get Started
These steps provide you with the information necessary to safely power-on the signal
generator and set up the front panel for local operation. The front panel of the signal
generator is your display's front panel. When you use the HP 70340A/41A in local operation,
its front panel menus and functions appear on the display screen.
1. Install the HP 70340A according to chapter 1 of the
applicable, the HP 70341A according to the
2. Check that the LINE VOLTAGE SELECTION switches of your mainframes/displays are
set to the voltage of your power source.
The LINE VOLTAGE SELECTOR switches are located as follows:
a. On the bottom of the HP 70001A mainframe
b. On the left side of the HP 70004A stand-alone display
c. On the rear panel of the HP 70205A or HP 70206A display
3. Set the system LINE (Mains) switches to OFF.
4. Check that the power cable is undamaged, and then connect it to the system and power
source.
Warning
This is a safety Class 1 product (that is, it is provided with a protective earth
terminal). An uninterruptible safety earth ground must be provided through the
power cable. Whenever it is likely that the ground is impaired, the product must
be made inoperative.
5. Set the mainframe/display LINE (Mains) switches to ON.
When you turn the MMS on, the power lights come on and the ventilation fans start. It
is normal for the HP 70340A/41A LEDs (including the ERR LEDs) to turn on during a
power-up test. The ERR LEDs should be o , however, when the power-up test is complete.
HP 70340A/41A User's Guide
HP 70341A Installation Guide
The HP 70340A/41A Quick Start Guide
1
and, if
.
1-1