High-Voltage Output
( o p t i o n 002)

Attenuator Test

This procedure continues from the previous one.
1. Set the DC offset to the positive and negative
voltages below. The digital voltmeter reading
be within the tolerances shown for each voltage.
DC Ofbet
f1.499 v
f499.9 mV
f149.9 mV
f14.99 mV
f4.999 mV f0.004929 to 0.005069 V
f1.499 mV f0.001464 to 0.001534 V
This procedure continues from the previous one.
the HP 3324A output directly to the voltmeter input.
2. Select the high-voltage output on the HP 3324A.
"HIGH-VOLT" will be shown in the display.
3. Set the HP 3324A DC offset to 20 V. The voltmeter
reading should be +19.775 to 20.225 V.
Tolerances
f1.48399 to 1.51401 v
f0.14838 to 0.15142 V
f0.04947 to 0.05051 V
f0.01482 to 0.01516 V
Performance Tests 12-31