HP 3324A Operating And Programming Manual page 185

Synthesized function/sweep generator
Hide thumbs Also See for 3324A:
Table of Contents

Advertisement

2. Set the HP 3324A output as follows:
3. Set the voltmeter as follows:
Note
The HP 3437A triggers on the negative going edge of the
HP 3324A sync squarewave.
4. Set the voltmeter delay to 0.00003 (seconds). Record
5. Measure the voltage at each 10% segment point by
High-Voltage Output Off
Function
Frequency
Amplitude
DC Offset
Range
Number of readings
Trigger
the voltmeter reading on the Performance Test
Positive Slope Measurement, (lO%)yl.
Record under
This is the 10% point on the positive slope of the
triangle (see Figure 12-9).
setting the voltmeter delay to the following. Enter
on the Performance Test Record in the appropriate
Positive Slope Measurement.
spaces under
Delay
0.000035
0.00004
0.000045
0.00005
0.000055
0.00006
0.000065
0.00007
Triangle
10 KHz
o v
1
External
Performance Tests 12-35