Required Equipment; With Option 010; Ohm Feedthru Termination -Hp- Model 11094B; 30 Mhz Reference Oscillator Frequency - HP 3336A Operating Manual

Synthesizer/level generator
Hide thumbs Also See for 3336A:
Table of Contents

Advertisement

Adjustments
of
Model 3336A/B/C
;
;
2
Required Equipment:
iv
©
dc Voltmeter
-hp- Model 3466A
a. Measure the dc voltage at the — 15V test point on the A2 Power Supply printed circuit
board.
b. Adjust A2R22 until the voltage is —15V, +0.03V.
c. Measure the dc voltage at the +5V and + 15V test points. These voltages should be
+5V, +0.05V and +0.03V, respectively.
d. If necessary, readjust A2R22 until all three voltages are within tolerance.
5-7. 30 MHz Reference Oscillator Frequency Adjustment.
5-8. This procedure sets the frequency accuracy of the 30 MHz Reference Oscillator to
within 5 parts per million.
Required Equipment:
Electronic Counter
-hp- Model 5328A
with Option 010
75 ohm Feedthru Termination
-hp- Model 11094B
e. If necessary, reconnect the special adapter removed in Step a.
5-9. Option 004, High Stability Reference Frequency Accuracy.
5-10. This
procedure
sets
the frequency
accuracy
of the High
Stability
Frequency
Reference, Option 004, to within 1 part in 10 million.
Required Equipment:
Electronic Counter
-hp- Model 5328A
&
with Option 010
75 ohm Feedthru Termination
-hp- Model 11094B

Advertisement

Table of Contents
loading

This manual is also suitable for:

3336b3336c

Table of Contents