Setting up a
Multi-Marker Interval
SINGLE
INTERVRL
OF 3
The marker interval can be swept either singularly or
continuously. A single sweep means that the HP 3324A
sweeps from the start frequency of the interval to the
stop frequency, and then remains there. A continuous
sweep means that the HP 3324A sweeps from the start
frequency of the interval to the stop frequency, and then
remains there for a programmed time (retrace time),
before starting again at the start frequency. The retrace
time ranges from 0 to 100,000 s. If a retrace time of 0
to 5 ms is entered, the retrace is automatic. This means
that the retrace time is set to the sweep time of the
interval.
This section will show how to set up a multi-marker
interval. Press lsweep to show the Sweep menu. To
choose the multi-marker mode, put the cursor on
"MARKER" and press (Select). A display similar to that
in Figure 7-10 will be shown.
1 1 3
1
LIY
Figure 7-10. The Multi-Marker Display
Sweep 7-17