The program string error bit is set when a command
string is received that the interpreter does not
understand, or when a command is not executable for
When the "IER" query
some reason.
number will be returned, with the following meanings:
Error
Number
Entry parameter out of bounds
Invalid delimiter
Frequency too high for waveform function
Sweep time too small or too large
Offset - amplitude incompatible
Sweep frequency too large for function,
start frequency too small,
sweep bandwidth too small,
start frequency greater than stop frequency
Unrecognisable mnemonic received
7
Unrecognisable data character received
8
High-voltage option does not exist
9
Index out of range for Xxx mnemonic
1 0
1 1
Missing comma
1 2
Numeric parameter out of range
1 3
Interval error in sweep sequence
1 4
Sweep sequence is too long
is sent, an error
Error