High-Voltage Output (Option 002) - HP 3324A Operating And Programming Manual

Synthesized function/sweep generator
Hide thumbs Also See for 3324A:
Table of Contents

Advertisement

High-Voltage Output
(Option 002) 1. c
12-6 Performance Tests
5. Set the HP 3324A to the following frequencies and
verify the specified levels, relative to the fundamental.
14.9 MHz
2 MHz
1.99 MHz
200 KHz
6. Set the HP 3324A frequency to 50 KHz and the
amplitude to 9.99 mV(p-p).
7. Set the spectrum analyzer controls to display the
fundamental and at least three harmonics. (It may be
necessary to decrease the video bandwidth to separate
the harmonics from the noise floor.) Verify that all
harmonics are at least 60 dB below the fundamental.
8. Set the HP 3324A to the following frequencies
and verify that all harmonics are 60 dB below the
fundamental.
10 KHz
1 KHz
100
Hz
Continued from the previous procedure.
high-impedance input as shown in Figure 12-1.
- 3 0 dB
- 3 0 dB
- 4 0 dB
- 4 0 dB