Initial Inspection

Warning
Power
Requirements
Caution
C-2 Installation
Inspect the shipping container for damage. If the
container or cushioning is damaged, it should be kept
until the contents of the shipment have been checked for
completeness and the instrument has been verified both
mechanically and electrically.
Procedures for checking the operation of the instrument
are given in Chapter 12 Performance Tests. If the
contents are incomplete, mechanical damage or defect
is apparent, or if an instrument does not pass the
operator's checks, notify the nearest Hewlett-Packard
office. Keep the shipping materials for carrier's
inspection. The HP office will arrange for repair or
replacement without awaiting settlement.
To avoid hazardous electrical shock, do not perform
electrical tests when there are signs of shipping damage
to any portion of the outer enclosure (covers, panels,
etc.).
Before applying AC line power to the HP 3324A, ensure
that the voltage selector on the HP 3324A bottom panel
is set for the proper line voltage and the correct line fuse
is installed in the fuse holder. Procedures for changing
the line voltage selector and fuse are contained in the
following section "Line Voltage Selection".
The HP 3324A can operate from any single-phase AC
power source supplying 100 V, 120 V, 220 V or 240 V in
the frequency range from 48 to 66 Hz (see Table C-l).