HP 3324A Operating And Programming Manual page 169

Synthesized function/sweep generator
Hide thumbs Also See for 3324A:
Table of Contents

Advertisement

5. Change the HP 3324A amplitude to 1.061 Vrms (3
6. Change the HP 3324A amplitude to 0.3536 Vrms
7. Function Test. Set up the equipment as shown in
HP 3324A
I
I
Sync
Figure 12-5. Function Amplitude Accuracy Test Set-Up
8. Set the HP 3324A as follows:
KHz and 100 KHz as above. Verify that all three
voltmeter readings are between 1.037 and 1.085
Vrms (ho.2 dB).
(1 V(p-p)) and set the DC offset to 1 mV. Set the
HP 3324A frequency to 100 Hz, 1 KHz and 100
KHz, and read the AC voltage. Verify that all three
readings are between 0.3370 and 0.3702 Vrms (f0.4
Figure 12-5.
Delayed
Oscilloscope
I
External
High-Voltage Option Off
Function
Frequency
Amplitude
DC Offset
DVM
External
50 Ohm
Square
99.9 Hz
o v
Performance Tests 12-19