5. Change the HP 3324A amplitude to 1.061 Vrms (3
6. Change the HP 3324A amplitude to 0.3536 Vrms
7. Function Test. Set up the equipment as shown in
HP 3324A
I
I
Sync
Figure 12-5. Function Amplitude Accuracy Test Set-Up
8. Set the HP 3324A as follows:
KHz and 100 KHz as above. Verify that all three
voltmeter readings are between 1.037 and 1.085
Vrms (ho.2 dB).
(1 V(p-p)) and set the DC offset to 1 mV. Set the
HP 3324A frequency to 100 Hz, 1 KHz and 100
KHz, and read the AC voltage. Verify that all three
readings are between 0.3370 and 0.3702 Vrms (f0.4
Figure 12-5.
Delayed
Oscilloscope
I
External
High-Voltage Option Off
Function
Frequency
Amplitude
DC Offset
DVM
External
50 Ohm
Square
99.9 Hz
o v
Performance Tests 12-19