2. Set the HP 3324A output as follows:
High-Voltage Output Off
Function
Frequency
Amplitude
DC Offset
3. Press c-1. After amplitude calibration
(approximately 2 seconds) the voltmeter reading
should be +4.350 to +4.650 Vdc.
4. Change the HP 332412 DC offset to -4.5 V. The
voltmeter reading should be -4.350 to -4.650 Vdc.
5. .Change the HP 3324A frequency to 999.9 KHz. The
voltmeter reading should be -4.440 to -4.560 Vdc.
6. Change the HP 3324A DC offset to +4.5 V. The
voltmeter reading should be +4.440 to +4.560 Vdc.
7. Set the HP 3324A function to square. The voltmeter
reading should be -f-4.440 to i-4.560 Vdc.
8. Change the HP 3324A DC offset to -4.5 V. The
voltmeter reading should be -4.440 to -4.560 Vdc.
9. Change the HP 3324A frequency to 9.9999 MHz.
The voltmeter reading should be -4.350 to -4.650
Vdc.
10. Set the HP 3324A function to triangle, frequency to
9.9 KHz. The voltmeter reading should be -4.440 to
-4.560 Vdc.
11. Set the function to positive ramp. The voltmeter
should be -4.380 to -4.620 V.
Sine
21 MHz
Performance Tests 12-33