Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 Table of contents Table of contents 1 User instructions ....................... 2 Safety..........................2.1 Infection hazard ......................7 2.2 Improper use ........................ 7 2.3 Technical condition ......................7 2.4 Accessories and combination with other equipment............
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
Page 5
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 1 User instructions Serial number Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity)
Page 6
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 1 User instructions CAUTION In cases which – if not prevented – can lead to material damage. 6 / 28...
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 2 Safety | 2.4 Accessories and combination with other equipment 4 Only operate devices or components if they show no signs of damage on the outside.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 2 Safety | 2.6 Service and repair 2.6 Service and repair Repairs, servicing and safety checks may only be performed by trained service personnel. The following persons are authorised to do this: ▪...
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 3 Description of the product | 3.1 Intended use 3 Description of the product EXPERTmatic LUX contra-angle handpiece E25 L (Mat. no. 1.007.5550) EXPERTmatic Contra-angle handpiece E25 C (Mat. no. 1.007.5551) 3.1 Intended use...
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 3 Description of the product | 3.2 Technical Specifications ▪ to prevent contamination from the product 3.2 Technical Specifications Drive speed max. 40,000 rpm Speed transmission...
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 4 Startup and shut down | 4.1 Checking the amount of water 4 Startup and shut down WARNING Hazard from contaminated products. Infection hazard to the dentist and patient.
Page 13
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 4 Startup and shut down | 4.1 Checking the amount of water 13 / 28...
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING Rotating dental bur. Cuts, infection and burn injury. 4 Never push the press-button while the dental bur is rotating.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 5 Operation | 5.2 Removing the medical device 4 Lightly spray O-rings on the motor coupling with KaVo Spray. 4 Place the medical device on the motor coupling and lock it into place.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 5 Operation | 5.4 Removing the dental bur CAUTION Dental bur with damaged, worn or deformed shafts. Injury hazard, dental bur may fall out during treatment.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions CAUTION Overheating of the device.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Press the O-ring between your fingers to form a loop. 4 Push the O-ring to the front, and remove it.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use WARNING Hazard from contaminated products.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing 4 Remove the medical device from the spray attachment and let the cleanser act for 1 minute.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.3 Automated reprocessing CAUTION Never reprocess the medical device with chloride-containing products. Malfunction and material damage. 4 Reprocess in a washer disinfector or by hand only.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing CAUTION Never reprocess this medical device in an ultrasonic cleaner.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing 7.4.1 Servicing with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use, i.e.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging NOTE Close the chuck service mode. Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A, close the front flap and start the service procedure.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 7 Reprocessing steps in accordance with ISO 17664 | 7.7 Storage 7.7 Storage Reprocessed products must be stored appropriately protected from dust in a dry, dark and cool space with a low germ level.
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691...
Instructions for use EXPERTmatic LUX E25 L - 1.007.5550, EXPERTmatic E25 C - 1.007.5551 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...