6.10 Patient positioning controls

6.10.1 X-ray unit up / down

Instructions for use
NOTE
Never allow patients to press the positioning controls when they are
positioned in the X-ray unit.
NOTE
Pressing any of the positioning controls (button or joystick) switches the
patient positioning lights on. The lights automatically switch off after two
minutes. To switch them off earlier, press the positioning joystick.
Use the X-ray unit up and down buttons to adjust the X-ray unit to suit the
height of the patient.
The X-ray unit moves slowly at first, then faster.
NOTE
If for some reason either of the buttons gets stuck during operation, you can
stop the up / down movement by pressing any of the other control buttons
or the positioning joystick. This is a safety measure that guarantees that the
up / down movement can be stopped in an emergency.
NOTE
Be careful that the X-ray unit does not hit the ceiling when you press the
up button. The maximum height can be adjusted to suit offices with a low
ceiling. Contact your service technician for help.
NOTE
Make sure that there is no object under the X-ray unit when you press
the down button. If something is in danger of becoming trapped, release
the button immediately to stop the movement. Clear any obstruction before
pressing the button again.