Installation, connection and commissioning

3. Installation, connection and commissioning
The CP9xx touch panel is integrated into existing LAN structures, but can also be operated via a single PC..
Only qualified personnel are permitted to carry out the work necessary to ins-
tall, commission and run a device or system.
Mounting flush-mounted enclosure
The flush-mounted enclosures supplied are only suitable for mounting in hollow
walls. In drywall and stud constructions, the enclosures must be screwed to a
batten or stud construction.

3.1 Installation of the CP9xx device

The COMTRAXX® CP9xx devices are installed exclusively in the supplied and professionally pre-assembled
flush-mounting boxes.

3.1.1 Dimension diagram

COMTRAXX®
b

3.1.2 Flush-mounting box CP907

75
128
7
37
Tightening torque
The tightening torque for the fastening screws may be in a range between
0.8 - 1 Nm.
CP9xx_D00349_07_M_XXEN/01.2020
a
c
124
dimensions in mm
Abb. 3.1: Basic enclosure shape and dimensions CP907
Device dimensions (mm) ±1
CP 907
a
226
b
144
c
176
216
CP 915
CP924
505
680
350
580
386
545
212
13