Switches - Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Advertisement

Section 6
Controls
Control Panel
(continued)

Switches

6-2
Nanopure Life Science (UV/UF)
The control panel can be removed from the unit and remotely mounted.
Refer to "Installing the Control Panel in a Remote Location" in the
"Installation" section of this manual for more information about how to
remove the control panel from the unit.
Note The removable control panel was not designed to be repeatedly
removed from the unit, therefore, you may experience difficulty when
attempting to remove it. It is therefore recommended that the control
panel be permanently mounted in a remote location or remain intact with
the unit.
s
When the main power switch (on the lower back right side of the unit as
you face the unit) is ON, the switches on the control panel function as
follows:
START/STOP allows you to alternate the unit between the normal (recir-
culation) mode; when the display is showing purity, and the (Idle) mode.
STANDBY allows you to put the unit into standby, recirculating water for
10 minutes/hour. The display will read "Standby" during periods of inac-
tivity and "Recirculating" and the time remaining during the 10 minute
recirculation.
DISPENSE allows you automatically deliver water from the unit. Please
refer to"Automatic Dispensing" in the "User Settings" section of this
manual for information.
UP and DOWN ARROWS allow you to scroll between menu
options/items and values.
BACK allows you to return to the previous menu option/item. Please note
that use of the BACK switch from a selection screen will return you to
the previous menu option/item while maintaining the selection's value
upon entry.
ENTER allows you to activate a selected menu option/item and also func-
tions as "Yes" whenever an option with a question mark appears.
DISPENSE KNOB when the unit is in the recirculation mode, push to
the right to deliver water continuously until pushed back to the middle.
Push to the left to deliver water manually until button is released.
Note The unit MUST be in the normal recirculating mode (when the
display is showing purity) in order to dispense water.
s
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

D11931

Table of Contents