System Cleaning
Procedure (cont.)
Thermo Scientific
6. Open front (left side door) to expose cartridge pack. Remove solution
syringe from wrapper.
Figure 7-2. Injecting Cleaning Solution
7. Remove luer cap on system injection port, located on the right side of
the cartridge pack manifold by turning counterclockwise.
8. Remove syringe luer cap and attach syringe to the system luer fitting.
9. Slowly inject solution into system and remove syringe. Avoid injecting
air into the system.
10. Replace luer cap on injection port and close door.
11. Press the UP or DOWN arrow to select (YES) and then press ENTER.
Display will now read, "Reconnected Water (NO)."
12. Attach feed water supply removed in Step 4.
13. Press the UP or DOWN arrow to select (YES) and press ENTER.
Display will now read, "Unit Cleaning."
14. Unit may now be left unattended until display reads, "Cleaning
Complete: Press ENTER."
15. After cleaning is complete, the unit will return to the (Idle) mode.
Injection Port
Section 7
Inital Operation
Nanopure Life Science (UV/UF)
7-5