Fuse Replacement - Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Advertisement

UV Lamp
Replacement (cont.)

Fuse Replacement

Thermo Scientific
5. Insert the lamp halfway into the UV lamp chamber and plug UV cable
in, then fully insert the lamp into the chamber.
Figure 12-5. UV Lamp Installation
6. Make sure the UV lamp cable is tucked behind the chamber before
closing and relatching the door.
7. Reconnect the power cord. Turn unit ON.
8. Reset UV Timer according to "Resetting the UV Timer."
9. Operate normally.
1. Turn off the Nanopure and disconnect it from the power supply by
removing the power cord directly below the main power switch on
right side of unit. Locate the fuse drawer above the power cord
receptacle.
2. Pull out the fuse drawer located in the power entry module.
3. Remove old fuses and replace with fuses of the same type and rating.
See "Replacement Parts."
4. Replace fuse drawer.
5. Reattach the power cord.
6. Operate normally.
Section 12
Maintenance and Servicing
Nanopure Life Science (UV/UF)
12-9