Summary of Contents for Thermo Scientific Partisol 2025i
Page 1
Partisol 2025i Sequential Air Sampler/ Partisol 2025i-D Dichotomous Sequential Air Sampler Instruction Manual Part Number 110100-00 14Apr2015...
Page 3
WEEE Compliance This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol: Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them.
Page 5
Chapter 1 “Introduction” introduces the user to the advanced features ● of the Partisol 2025i and Partisol 2025i-D and describes the flow and sampling configurations of the system. It also discusses the flow control scheme used in the unit.
Page 6
“AK Protocol” provides a description of the AK protocol ● commands that can be used to remotely control a Partisol 2025i and 2025i-D using a host device such as a PC or a datalogger. Safety Review the following safety information carefully before using the Sampler.
Page 7
A properly grounded antistatic wrist strap must be worn while handling any internal component. If an antistatic wrist strap is not available, be sure to touch the instrument chassis before Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 9
AMTIC web site: http://www.epa.gov/ttn/amtic/pmqa.html, the QC coordinator at any EPA Regional Office, or from the Monitoring and Quality Assurance Group (MD- 14), U.S. EPA, Research Triangle Park, NC 27711. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 10
Chinook Engineering, LLC. VSCC and Very Sharp Cut Cyclone are trademarks of BGI, Inc. Some content of this manual is courtesy of BGI, Inc. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 11
Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 12
About This Manual Where to Get Help viii Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 16
Flow Calibration Setup ..............9-28 Flow/Flow2 ................. 9-28 Setpoint 1-5 ................9-29 Number of Cal Points ..............9-29 Reset Flow Cal ................9-29 FTS Constant M/B ..............9-30 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 17
Default Gateway ................. 9-49 Host Name ................. 9-49 Network Time Protocol Server ............ 9-50 USB Setup ..................9-50 File Format ................. 9-50 Date Format ................9-51 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual xiii...
Page 18
USB Port ..................9-69 Export Data Logs ................ 9-69 Export Audit History ..............9-70 Backup Configuration ..............9-71 Restore Configuration ..............9-71 Update Firmware ................ 9-72 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 19
Bypass Shuttle ................9-93 Enable I/O Board ................ 9-93 Display Pixel Test ................9-94 Service Mode ..................9-94 Diagnostics ..................9-94 Program Versions ................ 9-95 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 20
Virtual Impactor Reinstallation ............ 10-16 V-Seals Cleaning and Replacement ..........10-16 V-Seals Cleaning and Replacement in the Partisol 2025i ... 10-17 V-Seals Cleaning and Replacement in the Partisol 2025i-D ..10-21 Particle Trap Filter Exchange ............10-25 ...
Page 21
Instrument Identification Number ............B-1 Commands ..................B-2 Commands List ................B-3 Measurements ..................B-7 Diagnostics ..................B-10 Datalogging..................B-11 Keys/Display ..................B-18 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual xvii...
Page 22
Executing RPComm ............... D-5 Creating a New Connection ............D-7 Downloading Stored Data ............D-11 Setting the Storage Pointer Positions .......... D-11 Downloading Data ..............D-12 xviii Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 23
Multiple Instruments of the Same Model ........D-27 Multiple Instruments of Different Models ......... D-29 Connecting a Series 1400a and a Partisol 2025i-D ..... D-30 Connecting a Model 2000-FRM and a Partisol 2025i/2025i-D Sampler ..................D-36 ...
1-10 ● “Specifications” on page 1-15 ● Overview Using a choice of sample inlets, the Partisol 2025i Sequential Air Sampler (Figure 1–1) has the ability to collect PM , PM , PM , and TSP PM. Properly configured collocated samplers can be used to collect and determine...
The supply and storage magazines are supplied with covers to seal off filter cassettes when not used for sampling. Figure 1–1. Partisol 2025i and 2025i-D Sequential Air Sampler Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Introduction Overview Inlet System The Partisol 2025i can be configured to operate with a variety of inlet systems, including the U.S. EPA VSCC or WINS inlet-based reference PM method, and PM , PM (using a Cyclone inlet), PM and TSP (total suspended particulate matter) inlets.
Introduction Advanced Features Advanced This section lists some of the advanced features of the Partisol 2025i and 2025i-D: Features Partisol 2025i The following is a list of some of the advanced features of the 2025i: Features Automatic filter changing system with a capacity of up to 16 filters per ●...
● Sampler and includes all U.S. EPA-specified values such as error condition flagging, and average temperatures and pressures. Filter data records also include sampled volume in volumetric and standard terms, Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
● life vacuum pump. Data can be retrieved using Ethernet, serial, or USB. ● Partisol 2025i The Partisol 2025i flow schematic provides an overview of the unit’s basic flow and electronic connections (Figure 1–2). The schematic shows an inlet Configuration system consisting of a first-stage inlet that is followed by a down tube and cut device.
For sensitive speciation applications, every second-loaded filter can be a separator, forming a complete seal between filter cassettes contained in the storage magazine. The instrument must be configured for separators prior to use. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
The mass flow sensor in the Partisol 2025i is calibrated at a temperature of 0 °C and pressure of 1 Atmosphere (1013.2 millibars or 760 mmHg). The instrument uses the measured ambient temperature and pressure to sample at the correct volumetric flow rate.
Page 39
The current pressure (mmHg) as measured by the pressure transducer in the Sampler’s enclosure. The Partisol 2025i automatically determines the sampled volume in volumetric and standard m³ for each filter exposed, and stores this information internally for later viewing or downloading.
Introduction Partisol 2025i-D Configuration Partisol 2025i-D The Partisol 2025i-D Dichotomous Sequential Air Sampler consists of two sample flows directed through two separate 47 mm filters that are housed in Configuration the filter platform (Figure 1–4). Figure 1–4. Partisol 2025i-D Flow Schematic Two 47 mm filters are housed in filter cassettes that the user installs with other cassettes in two supply cassette magazines.
Page 41
(1.7 L/min) flow, while the fine particle fraction follows both the minor and major (15 L/min) flow paths. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 1-11...
Three formulas are required for the mass concentration ( g/m³) computation of fine PM (PM ), coarse PM (PM minus PM ), and PM For fine PM: For coarse PM: For PM (in volumetric terms): 1-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 43
846 g. What is the mass concentration of fine PM, coarse PM and PM For fine PM (using equation 1): For coarse PM (using equation 2): For PM (using equation 3): Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 1-13...
Page 44
760 mmHg pressure and 25 °C temperature. Flow volumes referenced internally by the Sampler to 0 °C are converted to standard conditions using the following computation: 1-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Introduction Specifications Specifications Table 1–1 lists the Partisol 2025i and 2025i-D performance specifications. Table 1–1. Specifications Safety/Electrical Meets the following safety and electrical designations: Designations CE: EN550011 Group 1, Class B (Emissions); EN55082-1 (Immunity); EN611010-1 (Safety) ETL: UL- and CSA-equivalent approval...
Standard Hardware Configurations Setting up the Use the following procedure to set up the Sampler. The following instructions apply to both the Partisol 2025i and Partisol 2025i-D Sampler Ambient Samplers. 1. When moving the instrument, be sure to lift the unit by using the two handles.
Insert the end of the sample tube, with the 5 cm (2 inches) machined section, into the bulkhead of the Sampler. Ensure that the tube is pushed into the enclosure through the final O-ring until it stops. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Locate the two screws on the left side of the enclosure. b. Remove the two screws. Be sure to retain the washers. This will expose two holes. c. Locate the ambient temperature probe assembly in the compilation package. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Sampler. Follow the instructions to install the straight down-tube in the Sampler as provided in “Converting a Partisol 2025i from a PM2.5 Sampler to a PM10 Sampler” on page For PM sampling with the Partisol 2025i, an appropriate PM device, or fractionator, must be installed.
Put this hardware together in accordance with Figure 2–7 and the “Sampler Stand Parts List” that follows. ▲ Figure 2–7. Assembly Information for the Sampler Stand Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 55
4. Fasten the right and left rails to the legs and top front and back rails with F, G and H hardware (Figure 2–7). 5. Tighten all hardware. 6. Place the Sampler onto the stand and secure using E, G, and H hardware (Figure 2–7). Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
10 °C, the pump will continue running until the ambient temperature exceeds 10 °C. After instrument start up when the device is not sampling, the sample pump initiates if the ambient 2-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Screen Contrast necessary to change the screen contrast if the instrument is operated at extreme temperatures or lighting conditions. Note The optimal contrast will change with changes in temperature. ▲ Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 2-11...
Page 58
In the Main Menu, choose Service > Screen Contrast. ● SCREEN CONTRAST: CURRENTLY: 50 % SET TO: 60 % ? CHANGE VALUE SAVE VALUE CTRST SSET INSTS STAT 2-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 59
“Installing the Metal Piston Friction Ring” on page ▲ Filter Handling and The Partisol 2025i can be configured to sample a wide range of ambient PM sizes, including TSP, PM , PM , and PM . It is important that the...
Pallflex TX40 (Teflon coated glass fiber) filters, quartz fiber filters and Teflon material for PM sampling. The Partisol 2025i-D is specifically designed for sampling PM and as such the use of 47-mm diameter, 2 m pore-size Teflon coarse filters must be used for U.S.
These filter cassettes also are referred to as having “serialized” screens. Close the filter cassette by snapping its top part onto its bottom section. Ensure that the top and bottom pieces of the cassette are pushed completely together. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
As you place each cassette in the magazine, press down to move the piston downward, to make room for the next cassette. Note the order of the filter cassette serial numbers as they are inserted in the magazine. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Note The following figures show a clear magazine for informational purposes, however the actual magazines are not clear. ▲ Figure 3–3. Inserting Filter Cassettes into a Filter Cassette Magazine Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
The capped filter cassette magazines should be inserted into the metal transport container (20-004997) (Figure 3–6). The filter cassette magazine’s caps must be secure during transport to avoid contaminating the filters. Collection filters should be transported carefully in the metal Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The sample tube connects to the low flow channel when the enclosure lid is latched. In the software screens on the Partisol 2025i-D, the high flow (15.0 l/min), channel, is shown as “Flow” and the filter in the high flow sampling position is “Filter,”...
Page 66
“low flow” magazines to be installed in the front sampling compartment at the 1.67 L/min flow rate position. When installing the filter cassette magazines into the Partisol 2025i-D, install the rear high flow magazines first and then install the low flow magazines into the front positions of the filter shuttle mechanism.
Figure 3–10). Storage filter cassette magazine(s) do not have an air pressure supply tube. Figure 3–9. Positioning the Pressurized Air Tube Toward the Air Connector on the Filter Cassette Magazine Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
3. Place the orange cap over the open end of the filter cassette magazine to protect its contents during transport. Place the capped magazine into a metal transport container (Figure 3–6). 3-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The filter cassette magazine is now ready to be loaded with unused filters. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 3-11...
2. Remove the 47 mm filter from the filter cassette and set the filter in its petri dish. The cassette can then be used to hold other filters once it has been cleaned. 3-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
3–14). The average mass reading is the final filter weight, W (g). 5. Return the filter to its petri dish, place the petri dish cover over it and store it for archival purposes. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 3-13...
V = the volume drawn through the filter, as obtained from the Sampler. For U.S. EPA PM , this must be volumetric m³. Other sampling standards require that the standard volume be used. The 3-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Filter Handling and Exchange Installing the Metal Piston Friction Ring Partisol 2025i and Partisol 2025i-D Samplers report volumes in both the actual or standard volume. For 24-hour PM concentration averages to be valid without adjustment for U.S. EPA reporting purposes, the total sampling time must be between 23 and 25 hours, and other requirements must also be met, as referenced in the U.S.
Figure 3–17. Supply Filter Cassette Magazine with Bottom Manifold Removed 6. Slide the piston down and out of the bottom of the filter cassette magazine until the groove on the piston is exposed (Figure 3–18). 3-16 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
10. Reinsert the pin into one of the small holes in the bottom of the magazine, through the hole in the manifold, and out the other small hole in the side of the magazine. Push the pin in until it reaches a stop. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 3-17...
Note Be sure to remove the metal piston friction ring from all supply filter cassette magazines when operating your Sampler in warmer weather conditions (above 0 °C). ▲ Figure 3–20. Reinstalling the Supply Filter Cassette Magazine into the Sampler 3-18 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Sampler. This section does not apply to the Partisol 2025i-D. When installing a cyclone or WINS impactor in a Partisol 2025i, ensure that the proper adapter is used. The adapter for the VSCC has a groove machined around the upper lip. The adapter for the WINS impactor or other cyclones does not have a machined groove.
Sampler equipped with a WINS impactor to a Sampler to a PM Sampler. Sampler 1. Remove the PM inlet from the sample tube by pulling straight upward. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Partisol 2025i Inlet Conversion Converting a Partisol 2025i from a PM2.5 Sampler to a PM10 Sampler 2. Unlatch and open the top cover of the Sampler, with the sample tube and PM inlet attached, to gain access to the VSCC or WINS...
Partisol 2025i Inlet Conversion Converting a Partisol 2025i from a PM10 Sampler to a PM2.5 Sampler and agency. The U.S. EPA requires that these values be 25 °C for temperature and the standard pressure be 760 mmHg. Refer to “Standard Temperature”...
Wait mode offers the user the choice of entering the Audit or Stop operating modes. Sampling Mode While in the Sampling mode, the Sampler is currently in a user-defined sampling interval. Except in the case of Advanced sampling with Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Partisol Operational Modes conditions, the unit will draws ambient air in through its sampling system at the volumetric flow rates specified by the user. For the Partisol 2025i the default flow rate is 16.67 L/min and for the Partisol 2025i-D the default flow is 1.67 L/min for the coarse flow channel and 15 L/min for the fine...
Calibration menu, scroll to Audit mode and press . At the Audit Mode screen, press to exit the Audit mode, and press confirm the change. Press to resume regular operations. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
If the hardware is not in the Sampling Operating Mode at the time of ● the power failure, the unit resumes operation in its pre-outage operating mode upon restoration of power. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
), low flow (1.67 L/min) channel located in the front sampling coarse compartment. In the software screens on the Partisol 2025i-D, the high flow (15.0 L/min), PM channel, is shown as “Flow” and the filter in the high flow sampling position is “Filter,” and the low flow (1.67 L/min), PMc channel, is “Flow 2”...
Start Time 1 - The Start Time 1 parameter defines the time and date that the Sampler will begin sampling for the desired sample filter. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The permissible values of the Condition parameter are as follows: ------ Condition not used TEMP Current ambient temperature (°C) PRES Current ambient pressure (mmHg) %RH Current relative humidity (%) WNDSPD Current wind speed (km/h) WNDDIR Current wind direction (deg) Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Sampler itself is given a duration during which the sampling on an undetermined number of filters will take place. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The default flow rate parameter determines the sample flow rate (volumetric L/min). For the Partisol 2025i Sampler, the flow rate is typically 16.7 L/min. The Partisol 2025i-D Sampler uses two flow controllers to maintain the flow through the fine and coarse channels of the sampler. For the Sampler...
760. The default setting for Standard Pressure does not have any effect on the volumetric flow rate and actual volume calculations by the unit. This value is used to calculate standard sample volumes. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Current date and time parameter are the current local date and time (or other standard time selected by the user) expressed by default as hh:mm and the current local date expressed by default as yyyy/mmm/dd. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
8 for EPA verification schedule requirements. The verification procedures are similar for the two types of Partisol Samplers, the Partisol 2025i and Partisol 2025i-D. These procedures are described so they can be used by operators of both Samplers. In specific instances where the procedures need to differentiate between the 2025i and the 2025i-D, the information will be specifically indicated.
Press to return to the Audit and Calibration menu. 4. Remove the currently installed storage magazine(s) from the right side of the Sampler and cover with an orange storage cap. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
5. Remove supply magazine(s) from the left side of the filter shuttle mechanism and install on storage side. On the Partisol 2025i-D Sampler, ensure that the rear supply magazine is installed in the rear position on the storage side and the front supply magazine is installed in the front.
Audit screen. If the measured and Filter Temp readings are not within ± 2 °C [°C = 5/9 x (°F - 32)] perform the filter temperature calibration procedure. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
If the temperatures measured with the external thermometer and the recorded filter temperatures (Filter Temp and Filter2 Temp) are not within ± 2 °C [°C = 5/9 x (°F - 32)] perform the filter temperature calibration procedure. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
1. Determine the current ambient relative humidity (%). 2. Verify that the value for Ambient RH in the Audit screen is within ±1.5 percentage points of the measured ambient relative humidity. If Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
4. A Pass or Fail message will display at the end of the leak check cycle. The leak check pass criterion is a pressure drop of 25 mmHg or less. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Sampler (which was used to perform the external leak check) remains in the unit for the flow verification. ▲ While the flow verification procedures are similar for the Partisol 2025i and Partisol 2025i-D, since the Partisol 2025i-D Sampler uses two flow sensors to control the sample flows for the PM...
Sampler Operation Sampler Performance Verification Flow Audit/Verification – Use the following procedure to perform a flow audit on the Partisol 2025i: Partisol 2025i 1. Remove the flow audit adapter used during the leak check and install the Streamline Pro on the sample tube (Figure 7–7).
Use the following procedure to perform a flow audit on the 2025i-D: Partisol 2025i-D The audit procedure is similar to that of the Partisol 2025i single sample version of the Sampler. The main steps of the audit are the same, but must...
2. Scroll to Leak Checks, press , with the cursor at Internal, press The following message displays: MAKE SURE THAT A LEAK CHECK DISK IS IN THE SAMPLE FILTER LOCATION!! Press to begin testing. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 7-11...
Install the prepared filter cassette magazines in the filter shuttle mechanisms on the left side of the Partisol 2025i Samplers. A clean, empty storage magazine should be installed on the right side of each filter exchange mechanism.
Page 107
Apply Default Times which automatically applies the default time ● settings. To apply EPA default conditions (midnight-to-midnight, continuous, 24- hour sampling) scroll to Apply EPA Times in the Sample Setup menu, Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 7-13...
Page 108
Use the pushbuttons as indicated on the screen to set the Cassette2 ID number, and press to save the setting. Press to return to the Sample 1 menu. 7-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
4. Select the desired data records to download and follow the instructions on the instrument display. When complete, remove the USB stick from the Sampler. 5. Press repeatedly to return to the Main Menu. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 7-15...
, with the cursor at Filter, press , to display the filter records. Filter Record 1 is the oldest record. Filter Record 2 is the most recent record of the 32 maximum filters. 7-16 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Status soft key to display any status errors that are currently set. The Status screen displays current staus conditions, but does not display status conditions from previous completed samples. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 7-17...
For these Calibration procedures, prepare a set of audit magazines consisting of a supply magazine and a clean empty storage magazine. For the Partisol 2025i-D Sampler, two sets of magazines are required. Audit magazine requirements are described in detail in the Audit and Verification section above.
Page 114
4. On the Partisol 2025i-D Sampler, ensure that the rear supply magazine is installed in the rear position on the storage side and the front supply magazine is installed in the front.
To convert from Atmospheres @ 0 °C to mmHg, multiply by 760. To convert from millibars to mmHg, multiply by 0.75012. To convert from inches Hg @ 32 °F to mmHg, multiply by 25.4. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
3. Enter the measured ambient pressure and press . The Sampler automatically adjusts the corresponding offset based upon this input. 4. If the Sampler is a Partisol 2025i-D, repeat these steps to calibrate the filter2 pressure sensor. Filter Compartment Use the following procedure to calibrate the filter compartment temperature.
°C [°C = 5/9 x (°F - 32)] and press save the value. The Sampler automatically adjusts the corresponding offset based upon this input. 5. If the Sampler is a Partisol 2025i-D, Filter2 Temp must also be calibrated. 6. Press...
Enter the m and b constants for the FTS sensor(s) into the Sampler. If the Sampler is a Partisol 2025i, enter a single set of constants. If the Sampler is a 2025i-D, enter constants for two flow sensors. To use the FTS as the calibration device, use the following procedure.
Use FTS Constants is set to Yes as described above. 4. Press repeatedly to return to the Main Menu. Scroll to Audit and Calibration and press to display the Audit and Calibration Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 120
Thermo recommends that these setpoints be used as they are configured for 10% above and below the usual setpoint. For the Partisol 2025i Sampler, the following values are recommended for a three-point calibration of the 16.7 L/min flow rate setpoint: Setpoint 1 (Min Flow): 15.0 L/min (10% below the usual 16.7...
2. While in Flow2 Cal Setup screen verify the number of points and set points meet your elected settings. Verify the setup for the high flow channel or ‘Flow’ also matches your elected settings. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 122
6. From the Calibration menu select ‘Advance Filter’ to advance the cassettes into the sampling position. 7. Scroll to ‘Flow’ and press enter a. To start the calibration press enter with the cursor in front of ‘Start Calibration’ 8-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 123
8. Using the ‘menu’ key escape back to the Calibration menu. Place any two cassettes in the supply magazines (left side) of the filter shuttle Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 8-11...
Page 124
Install the NIST-traceable volumetric flow meter onto the sample inlet tube after removing the PM10 inlet. Allow about 2 minutes for the flow meter reading and the unit ‘Live measured’ value to stabilize 8-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 125
To audit both flows or the ‘Total Flow’ @ 16.7 LPM the ‘slotted’ calibration/audit cassettes is not used and should be replaced with another leak free regular cassette with filter media installed. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 8-13...
(Figure 9–1). Some menus contain more items than can be displayed at one time. For these menus, use to move the cursor up and down to each item. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
If the liquid crystal contacts your skin or clothes, wash it off immediately using soap and water. ▲ Pushbuttons The Pushbuttons allow the user to traverse the various screens/menus (Figure 9–2). Figure 9–2. Front Panel Pushbuttons Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
“edit soft key prompt” screen will not come up upon entering right-arrow-soft key combinations. All items under the Service menu (including the menu itself) cannot be assigned soft keys. ▲ Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
You can customize the appearance and contents of the Run screen. See “Run Screens” on page 9-75. From the Run screen, the Main Menu can be displayed by pressing Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
This chapter describes each submenu and screen in detail. Refer to the appropriate sections for more information. Figure 9–3. Partisol 2025i and 2025i-D Flowchart of Menu-Driven Firmware Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
The word “STOP” on the left of the status bar indicates the Sampler is in “STOP” mode. Other modes appear in the same area of the display. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
▲ to move the cursor up and down. ● Press to make a selection. ● Press to return to the Main Menu or to return to the Run ● screen. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
SAMPLE02 2010AUG25 23:00 CTRST SSET INSTS STAT SAMPLE16 2010AUG25 23:00 Auto Populate Auto Populate populates the sample list and calculates the default start time and date once sample 01 is programmed. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Note that the Repeat Time is not shown directly, it is implied by the difference in Start time from sample to sample. In the Main Menu, choose Sample Setup > Sample (1-16). ● Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
If you set this parameter to Yes, the first filter will be a blank filter and the next filter advanced by the Sampler will be the sampling filter 9-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
ID numbers are entered, incremented and edited in the manner described above for Filter ID numbers. The Cassette2 ID screen is used to enter the cassette number for the 2025i- Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-11...
Interval Data - The Sampler writes a new record of interval data every 5 minutes. Each record contains the latest 5-minute average of the filter temperature, ambient temperature, ambient pressure and average flow rate. 9-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
-- DATE -- TIME CTRST SSET INSTS STAT FILTER REC. STATUS Status: OK CTRST SSET INSTS STAT Format of Filter Data Each filter data record contains: Records Filter ID (32-character string) 9-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 141
Minimum Ambient RH (%) Average Ambient RH (%) Maximum Ambient RH (%) Valid Sampling Time (mmmm) Total Sampling Time (mmmm) Filter Identification 2* Cassette Identification 2* Flow Minimum 2 (L/min)* Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-15...
The Sampler has the capacity to retain more than 17 days of input data (at one minute logging of 18 data items) before it overwrites the oldest records. 9-16 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
5 minutes. Each record contains the latest 5-minute average of the following: Ambient Temp Filter Temp Ambient Pres Flow Filter2 Temp* Flow2* 2025i-D only Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-17...
0000000 TOTAL USERS: 1574 MOVE CURSOR CHANGE VALUE SAVE CTRST SSET INSTS STAT to input the set back number as appropriate and press to save the records. 9-18 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
SET CURSOR TO MONTHS ACCEPT AS SHOWN CTRST SSET INSTS STAT to select the date and time. Use to change the date and time, and press to display the records. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-19...
: Flow = 15.00 L/min; Flow2 = 1.67 L/min IMPORTANT Flow2 Rate and Filter2 Type apply only to the Partisol 2025i-D. ▲ In the Main Menu, choose Instrument Setup > Default Sample Setup. ● 9-20 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Filter Type/Filter2 Type The Filter Type screen allows the user to identify the filter type. The Filter2 Type screen (2025i-D only) is used to identify the filter type for the 2025i-D. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-21...
The repeat time is reflected in the start time of the next filter. Changing the start time will cause the start time (and stop time) of all the following filters to be updated. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-23...
The Signal screen allows the user to select signal for the selected trigger. In the Main Menu, choose Instrument Setup > Advanced Trigger > ● Setup Trigger 1-3 > Signal. 9-24 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
AUTO RUN Average Temperature The Average Temperature screen is used to adjust the average temperature. The default for this parameter is 99 °C indicating that the unit should use Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-25...
In the Main Menu, choose Instrument Setup > System Setup > Filter ● Fan. FILTER FAN MODE: CURRENTLY: AUTOMATIC SET TO: AUTOMATIC CHANGE VALUE SAVE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-27...
STAT Flow/Flow2 The Flow Cal Setup menu presents the parameters used to setup flow calibration. In the Main Menu, choose Instrument Setup > Flow Calibration Setup ● > Flow/Flow2. 9-28 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Reset Flow Cal menu item is used to reset the flow calibration parameters. In the Main Menu, choose Instrument Setup > Flow Calibration Setup ● > Flow > Reset Flow Cal > press to reset the flow calibration parameters. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-29...
The selected item is shown by “<--” after it. Note that at this point, pressing indicates that these are proposed changes as opposed to implemented changes. To Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-31...
● Select Content > select field > Analog Inputs. ANALOG INPUTS: >NONE ANALOG IN 1 ANALOG IN 2 ANALOG IN 3 ANALOG IN 4 CTRST SSET INSTS STAT 9-32 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 159
ELEC TEMP TEMP DLT TEMP2 DLT TEMP DLT MAX TEMP2 DLT MX RELATIVE HUM WIND SPEED WIND DIR WIND SPD AVG WIND VEL WIND VEL AVG WIND DIR AVG Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-33...
TO CONFIRM RESETxx CTRST SSET INSTS STAT CTRST SSET INSTS STAT Configure Datalogging The Configure Datalogging menu is used to configure the datalogging for the currently selected record type. 9-34 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Configure Datalogging > Memory Allocation Percent. SET PERCENT USERS: CURRENTLY: SET TO: THIS WILL ERASE ALL LOGS! PRESS TO CONFIRM CHANGE CHANGE VALUE SAVE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-35...
Page 162
“Erasing – Please Wait” message to confirm the erasure. In the Main Menu, choose Instrument Setup > Datalogging > Erase ● User Data Log. 9-36 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Sampler’s data output from this screen. The Site ID2 screen items are used in the same manner as the Site ID1 screen items. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-37...
The Sampler’s default baud rate is set to 9600 to provide backwards compatibility with the older C-series analyzers. In the Main Menu, choose Instrument Setup > Communications > Serial Settings > Baud Rate. 9-38 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Selections of 1 and 2 are available, with 1 being the default value. In the Main Menu, choose Instrument Setup > Communications > Serial Settings > Stop Bits. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-39...
Valid instrument ID numbers are from 0 to 127. In the Main Menu, choose Instrument Setup > Communications > Instrument ID. 9-40 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The MODBUS protocol support for the Partisol 2025i and 2025i-D enables the user to read analog values or variables, read the status of the digital outputs of the instrument, and to trigger or simulate the activation of a digital input to the instrument.
No. If set to Yes, then each streaming data record will start with a time and date stamp indicating when that record was generated. In the Main Menu, choose Service > Communications > Streaming Data Config > Prepend Timestamp. 9-42 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
(none or analog inputs 1-8) that will be included in the streaming record. If the I/O expansion board is not present, Analog Inputs will not be available. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-45...
FILTER ID FILTER2 ID CASS ID CASS2 ID CTRST SSET INSTS STAT START DATE STOP DATE ELAPSED TIME FS STEP FS STATUS TIME DATE DATE/TIME EXT ALARMS STATUS CODES 9-46 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Note The instrument power must be cycled after any of these parameters have been changed for the change to take effect. ▲ In the Main Menu, choose Instrument Setup > Communications > TCP/IP Settings. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-47...
In the Main Menu, choose Instrument Setup > Communications > TCP/IP Settings > IP Address. IP ADDRESS: CURRENT: 192.168.1.200 SET TO: 192.168.001.015 MOVE CURSOR CHANGE VALUE SAVE VALUE CTRST SSET INSTS STAT 9-48 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Host Name screen is used to edit the host name. When DHCP is enabled, this name is reported to the DHCP server. In the Main Menu, choose Instrument Setup > Communications > TCP/IP Settings > Host Name. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-49...
>FILE FORMAT DATE FORMAT 20110310 FORMAT USB DRIVE CTRST SSET INSTS STAT File Format The File format screen is used to select the format of the export data log files. 9-50 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 177
ERASES ALL FILES! ARE YOU SURE?XXX FORMATTING ... PRESS TO CONFIRM PRESS TO CANCEL ERASING - PLEASE WAIT ● CTRST SSET INSTS STAT CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-51...
CTRST SSET INSTS STAT Password Menu The Password menu allows the user to configure password protection. If the instrument is “locked,” none of the settings may be changed via the 9-52 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Instrument, the instrument reports being in Remote mode. In this mode, Local/Remote Operation the front panel is “locked,” data and settings can be viewed, but not changed using the front panel interface, and the remote “Set” commands are active. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-53...
In the Main Menu, choose Instrument Setup > Password > Change ● Password. ENTER NEW PASSWORD: RANGXXXXXE ABCDEFGHIJKLMN BKSP OPQRSTUVWXYZ PAGE 0123456789 ./- SAVE CTRST SSET INSTS STAT 9-54 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
In the Main Menu, choose Audit and Calibration. ● Menu AUDIT AND CALIBRATION: >AUDIT MODE ADVANCE FILTER AUDIT VIEW AUDIT HISTORY CALIBRATION CTRST SSET INSTS STAT LEAK CHECKS Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-55...
STAT View History The View History screen is used to view history. In the Main Menu, choose Audit and Calibration > Audit > Ambient ● Temperature > View History. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-57...
The Calibration menu displays calibration related menu items. Filter2 Pres and Flow2 are used with the 2025i-D only. In the Main Menu, choose Audit and Calibration > Calibration. ● 9-58 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
CHANGE VALUE SAVE CTRST SSET INSTS STAT Filter Temperature The Calibrate Filter Temperature screen displays the current averaged values of the filter #1 (high flow filter) temperature (°C). Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-59...
Flow2 menu items are used in the same manner as the Calibration for Flow menu items. In the Main Menu, choose Audit and Calibration > Calibration > ● Flow. 9-60 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Cal Setup The Cal Setup Flow screen provides parameters for flow calibration setup. In the Main Menu, choose Audit and Calibration > Calibration > Flow ● > Cal Setup. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-61...
Reset Flow Cal menu item is used to reset the flow calibration to the default value. In the Main Menu, choose Audit and Calibration > Calibration > Flow ● > Cal Setup > Reset Flow Cal. 9-62 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Leak Check menu provides access to the internal leak check parameters and the external leak check parameters used in performing leak checks. In the Main Menu, choose Audit and Calibration > Calibration > Leak ● Checks. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-63...
● Checks > External. LEAK CHECK EXTERNAL: MAKE SURE THAT FILTER IS IN PLACE AND LEAK CHECK ADAPTER IS CLOSED!!! PRESS TO BEGIN TESTING CTRST SSET INSTS STAT 9-64 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The unit stores all status conditions that apply to each exposed filter in the filter data storage buffer. In the Main Menu, choose System Status > Status Codes. ● Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-65...
Filter Temperature 1 Range. The measured temperature of Filter #1 (high flow filter) exceeded the measured ambient temperature by more than 5° C for more than 30 consecutive minutes. 9-66 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 193
Blank Filter. Indicates that “sample” was a zero time blank. Hi Wait Flow. This status condition indicates there is flow greater than 1 L/min through the system in Wait mode, when there should be no flow. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-67...
Service mode. Refer to the “Service Mode” on page 9-94 information on the Service mode. ▲ In the Main Menu, choose USB. ● USB: >FRONT USB PORT CTRST SSET INSTS STAT 9-68 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Export Data Logs screen indicates the status of the process and then displays the name of the file stored on the USB memory device. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-69...
Page 196
The Export Audit History screen is used to export an audit history to a USB memory device. In the Main Menu, choose USB > Front USB Port > Export Audit ● History. EXPORT AUDIT LOG: PROGRESS CTRST SSET INSTS STAT 9-70 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The menu displays the 32 most recent .tbf files in the 2025i directory on the USB memory device. In the Main Menu, choose USB > Front USB Port > Restore ● Configuration. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-71...
Note The Service mode must be On to access the Service Menu. Refer to “Service Mode” on page 9-94 for information on the Service mode. ▲ In the Main Menu, choose Service. ● 9-72 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
STATUS HALL SENSORS Discrete Actions The Discrete Actions allows the user to choose direct action menu. In the Main Menu, choose Service > Manual Motion > Discrete ● Actions. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-73...
Notes The optimal contrast will change with changes in temperature. ▲ The optimal contrast will change from one LCD screen to another. If the LCD screen is replaced, the contrast may need to be reset. ▲ 9-74 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Run screen. This menu allows the user to edit the Run screen title, toggle the Run screen On and Off, specify the number of items to be displayed on the Run screen (1-10), and to assign a parameter to each item. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-75...
Note You cannot scroll through Run screens when a menu screen is displayed. ▲ In the Main Menu, choose Service > Run Screens > Select a Run ● Screen > Enabled. 9-76 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
In the Main Menu, choose Service > Run Screens > Select a Run ● Screen > Item number. SELECT RUN SCREEN ITEM: >OTHER MEASUREMENTS ANALOG INPUTS NON-MEASUREMENTS OTHER MEASUREMENTS 2 CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-77...
Pressure in mmHg at the filter for the fine channel Filter2 Pressure Pressure in mmHg at the filter for the coarse channel Pump Pressure Pressure at the pump in atmospheres Ambient Pressure Average Average ambient pressure in mmHg 9-78 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The temperature of filter 1 minus the ambient temperature in C Temp Dlt Max The maximum temperature difference of filter 1 over the sample period in C Relative Hum The relative humidity in percent Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-79...
Run screen. Refer to Item Number described previously. In the Main Menu, choose Service > Run Screens > Select a Run ● Screen > Item number > Non-Measurements. 9-80 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
In the Main Menu, choose Service > External I/O > Output Relay ● Settings. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-81...
The actual use of these inputs will vary based on the application. Note The digital inputs must be asserted for at least one second for the action to be activated. ▲ Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-83...
4 mA. By default this parameter is set to on, and 5% over and under range is allowed for all analog output channels. In the Main Menu, choose Service > External I/O > Analog Output ● Configuration > Allow Over/Under Range. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-85...
Minimum and Maximum The Minimum Value screen is used to edit the zero (0) to full-scale (100) Value value in percentages for the selected analog output channel. See Table 9–9 9-86 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
800 mmHg Filter Pressure Average Filter2 Pressure Average Pump Pressure 0 atm 5 atm Pump Pressure Average Ambient Temperature Filter Temperature -60 °C 60 °C Filter2 Temperature Filter Compartment Temperature Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-87...
Page 214
In the Main Menu, choose Service > External I/O > Analog Output ● Configuration > Select Channel > Choose Signal To Output > Choose a signal type. CHOOSE SIGNAL TYPE: >CONCENTRATIONS OTHER MEASUREMENTS ANALOG INPUTS CTRST SSET INSTS STAT 9-88 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
2-10 table points, and corresponding number of points selected. In the Main Menu, choose Service > External I/O > Analog Input ● Configuration. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-89...
The units may be from 1 to 3 characters in length, and defaults to V (volts). In the Main Menu, choose Service > External I/O > Analog Input ● Configuration > select channel (1-8) > Units. 9-90 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
In the Main Menu, choose Service > External I/O > Analog Input ● Configuration > select channel (1-8) > Table Points. NUMBER OF TABLE POINTS: CURRENTLY: SET TO: 10 ? INC/DEC SAVE VALUE CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-91...
1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U, where U is the previously entered unit of measure. In the Main Menu, choose Service > External I/O > Analog Input Configuration > select channel > select point > User Value. 9-92 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
When an I/O expansion board is removed from the instrument, the Enable I/O Board menu item should be set to “No” to prevent an I/O board alarm from being triggered. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-93...
This menu is useful when troubleshooting the instrument. The analog input readings and analog input voltages are displayed only if the optional I/O expansion board is installed. In the Main Menu, choose Diagnostics. ● 9-94 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Motherboard Voltages The Motherboard Voltages screen (read only) is used to display the voltage readings on the motherboard. In the Main Menu, choose Diagnostics > Voltages > Motherboard. ● Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-95...
The Analog Input Readings screen (read only) displays the eight user-scaled analog input readings (if the I/O expansion board option is installed). In the Main Menu, choose Diagnostics > Analog Input Readings. ● 9-96 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Relay States screen displays the state of the ten digital outputs and eight solenoid outputs, and allows toggling of the state to either On (1) or Off (0). The relays are restored to their original states upon exiting this screen. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-97...
“normal” operating mode. In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL, ● Voltage Channel 1–6, or Current Channel 1–6. 9-98 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The Contact Information screen displays the customer service information. In the Main Menu, choose Diagnostics > Contact Information. ● CONTACT INFORMATION: CALL CENTER: 508-520-0430 WWW.THERMO.COM CTRST SSET INSTS STAT Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 9-99...
50 °C. Upper and Lower Cassette Seals - Inspect the seals that rest against the filter cassette every time a filter is exchanged. Wipe the seals with a clean Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-1...
Rebuild Kit (59-012071) (one kit is required for each rebuild), or replaced with a new pump. Partisol 2025i In addition to routine maintenance, see the following for Partisol 2025i: Maintenance Items VSCC Cyclone - Clean the VSCC every 30 days of sampling. The cyclone must be cleaned when the unit is not sampling.
2. Using a Phillips-blade screwdriver, remove the four pan head screws from the top of the top plate. Lift the top plate off the four, threaded, spacer standoffs and set aside. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-3...
Page 230
Dry and reinstall. 4. Using a general-purpose cleaner with a paper towel, clean the top plate deflector cone and internal wall surface of the acceleration assembly. 10-4 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
1- 1/4-inch OD sample tube. 4. Clean the lower collector assembly’s threads to ensure a tight seal when the two halves are reassembled. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-5...
● ● ● Only the PM VSCC is approved for use in a Partisol 2025i for official U.S. EPA PM sampling. Other Cyclones may be approved for PM ) sampling by other agencies. To use the Cyclone fractionator in EPA-designated reference and equipment Samplers, you must clean the fractionator every 30 days of operation.
1. Remove the cyclone from its installed position in the Sampler. The remove of the cyclone is described in the Sampler audit/verification section above. 2. Remove the Partisol 2025i Sampler adapter from the cyclone. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Figure 10–3, which shows the two adapters used with the WINS, VSCC, and SCC fractionators in the Partisol 2025i. The adapter used with the WINS impactor and SCC Cyclones (right side of the figure) has a smooth ungrooved surface around the top of the adapter. The adapter used with the VSCC (left side of the figure) has a groove machined into the top section of the adapter.
IMPORTANT If you are performing an audit from the Sampling mode and you choose Stop in the Audit Confirmation screen, the Sampler will not resume sampling (after your audit procedures are finished) at the set points Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-9...
Figure 10–3, which shows the two adapters used with the WINS, VSCC, and SCC fractionators in the Partisol 2025i. The adapter used with the WINS impactor and SCC Cyclones (left side of the figure) has a smooth ungrooved surface around the top of the adapter. The adapter used with the VSCC (right side of the figure) has a groove machined into the top section of the adapter.
Apply a thin coating of O-ring lubricant onto the O-ring, if necessary. 8. Place a new 37 mm borosilicate, glass-fiber filter (32-004294) onto the bottom of the impactor assembly. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-11...
2. Reinsert the WINS impactor into the upper section of the enclosure by pressing downward. 3. Lower the top cover and re-latch it. 4. Perform a system leak check. 10-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Audit mode. Virtual Impactor Use the following procedures to remove, clean, and reinstall the virtual impactor that is used in the Partisol 2025i-D Sampler. Maintenance Virtual impactor maintenance and performance verification can be performed while in the Stop, Wait, or Sampling modes. If the Sampler is in the Stop Mode, press and select Audit and Calibration >...
1. Unscrew the four screws on each corner of the bottom section of the virtual impactor. Separate its probe housing from its base plate (Figure 10–9). Base Plate Nozzle Housing Probe Houseing Figure 10–9. Virtual Impactor Sections 10-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
A general-purpose cleaner can be used, if necessary. Probe Collector Tube Figure 10–11. Interior View of the Virtual Impactor's Probe Housing Exhaust Tube Figure 10–12. Interior View of the Virtual Impactor's Base Plate Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-15...
> Audit and press to turn Off the Audit mode. V-Seals Cleaning As described above, the V-Seals in the Partisol 2025i and Partisol 2025i-D Samplers require periodic maintenance. The V-seals should be checked and Replacement every four weeks and replaced as necessary. In addition, anytime there is a suspected leak in the system that can be attributed to the V-seals, they should be checked and replaced if necessary.
Sampler enclosure (Figure 10–14). Note The device is installed with O-rings. Lift and slightly rock the device back and forth to remove it. ▲ Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-17...
Figure 10–15. Removing the Electronics Cover 5. Remove the side cover of the electronics compartment by removing the four small screws accessible from the main compartment (Figure 10– 16). 10-18 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
7. Move the plate (with the top head attached) to the side to expose the V-seal in the filter shuttle mechanism (Figure 10–18). Lower V Seal Figure 10–18. Filter Shuttle Mechanism (lower) V-Seal Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-19...
Figure 10–20. Looking Inside Top Cover 10. Replace the top head mounting plate and secure using the four screws. 11. Install the side cover, top cover, and exit cylinder cover. 10-20 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Partisol 2025i-D 1. Ensure the Sampler is not sampling and is in Stop mode. 2. Unlatch and open the top cover of the Partisol 2025i-D Sampler with the down tube and PM-10 inlet attached to gain access to the virtual...
Exit Cylinder Cover Electronics Cover Figure 10–23. Removing the Electronics Cover 5. Locate the front top head mounting plate and remove the four screws holding it in place (Figure 10–24). 10-22 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
10–26). The bottom seal is located in the now exposed section of the filter shuttle assembly. Examine both seals for drying and/or cracking or over compression. Replace, if necessary. Bottom V-seal (23-002182) Top V-seal (23-004276) Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-23...
Figure 10–27. Looking Inside Top Cover 11. Install the electronics compartment cover. 12. Replace the virtual impactor and close the top cover. 13. Perform a leak check. 14. Resume normal operation. 10-24 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
5. Remove the gasket and top disk from the filter stand. Note that the gasket sometimes remains inside the filter manifold. 6. Remove the filter element from the filter stand and install a new element. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-25...
O-ring on the piston, the large black seal encircling the manifold, or the small black O-ring on the quick disconnect hose barb (attached to the manifold). Obvious damage or signs of wear to the O- 10-26 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 253
O-rings and seals, you must reassemble the filter cassette magazine. Use the Reassembly following procedure to reassemble a filter cassette magazine. 1. Slide the piston into the bottom of the magazine tube. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 10-27...
Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 10-28 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Sampler’s current state. Table 11–1 lists the Diagnostics menu items and describes their function. Refer to the “Diagnostics” on page 9-94 for detailed information about the Diagnostics menu. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 11-1...
Guide information and indicates the checks that you should perform if you experience an instrument problem. For additional service assistance, see “Service Locations” on page 11-24. 11-2 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Check ribbon cable on side of display disconnected board. Board-Level Figure 11–1 Figure 11–2 can be used along with the connector pin descriptions in Table 11–3 through Table 11–7 to troubleshoot board-level Connection Diagram faults. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 11-3...
Page 265
KEYPAD_ROW3 KEYPAD_COL2 KEYPAD_COL1 KEYPAD_COL4 KEYPAD_COL3 +24V +24V AC IN NEUTRAL LIVE EARTH AC 24V POWER NEUTRAL SUPPLY LIVE EARTH AC INTF BOARD NEUTRAL LIVE EARTH DISPLAY LCDOUT_CLK LCDOUT_LP LCDOUT_FLM Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 11-11...
Analog Voltage Input 3 Ground Analog Voltage Input 4 Analog Voltage Input 5 Analog Voltage Input 6 Ground Analog Voltage Input 7 Analog Voltage Input 8 Ground Current Output Return Ground Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 11-23...
Page 278
Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 11-24 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 279
When the instrument is unplugged, the chassis is not at earth ground. ▲ Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 12-1...
3. Insert the USB memory device with the firmware update into the USB port on the Sampler. 4. Scroll to USB on the Main Menu and press 5. Select the USB port and press 12-2 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Thermo Fisher Scientific warranty. You may purchase consumable items from Thermo Fisher Scientific by calling our customer service department at 1-866-282-0430. ▲ Table 12–1. Partisol 2025i and 2025i-D Replacement Parts Replacement Parts Part Number Electronics Enclosure w/Processor PCB, 120VAC*...
6-Pin Cable Connector 06-002454 25-Pin Wiring Adapter 06-004521-0025 Internal Cables Table 12–2 lists the internal cables for the Partisol 2025i and 2025i-D. Table 12–2. Partisol 2025i and 2025i-D Internal Cables Part Number Description 101035-00 Measurement Interface Board to 24V Power Supply...
Page 286
AC Switch to Motherboard 110193-00 IEC to Ground (2 required per instrument) 110194-00 Measurement Interface Board to Modem IEC 110288-00 I/O Expansion Board Patch 110289-00 RS-232 Patch 110291-00 USB Patch 12-8 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
● one snap track ● Note Supporting all of the connections on units with the optional I/O expansion board requires: ▲ two DB37 kits ● one DB25 kit ● Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 12-9...
This filter log can be used to keep track of all important readings associated with each exposed filter. Users are encouraged to make photocopies of the form on the following page or to use a similar format. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 12-11...
Toll Free U.S. only 1-866-282-0430 U.S., Latin America, and Canada 1-508-520-0430 Europe +31 76 579 5555 China +86 10 8419 3588 Asia Pacific +91 22 27781102 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 12-13...
Analog input (voltage) ● Digital output (TTL levels) ● Digital input (TTL levels) ● Note The sampler has spare solenoid valve drivers and I/O support for future expansion. ▲ 13-2 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Only two points are necessary for linear inputs, however a larger number of points may be used to approximate non-linear inputs. All voltage inputs have a resolution of 12 bits over the range of 0 to 10 volts. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 13-3...
(set using a DHCP server). Up to three simultaneous connections are allowed per protocol. Any serial port protocols may be accessed over Ethernet in addition to the serial port. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual 13-5...
The USB port is USB 2.0, full speed, 12Mbps; and low speed, 1.5Mbps and is used for interacting with a USB memory device. The USB memory device must be formatted as FAT32. 4 gB or greater is recommended. 13-6 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
(i) normal wear and tear, (ii) accident, disaster or event of force majeure, (iii) misuse, fault or negligence of or by Buyer, (iv) use of the Products in a manner for which Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 300
OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
C-Link Protocol Commands This appendix provides a description of the C-Link protocol commands that can be used to remotely control the Partisol 2025i and 2025i-D Samplers using a host device such as a PC or a datalogger. C-Link protocol may be used over RS-232, RS-485, or Ethernet. C-Link functions can be accessed over Ethernet using TCP/IP port 9880.
It is important that each time instrument parameters are changed, that this command be sent. If changes are not saved, they will be lost in the event of a power failure. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Commands List Table B–1 lists the Partisol 2025i and 2025i-D Sampler C-Link protocol commands. The interface will respond to the associated command strings. Items marked with an “*” apply to the Partisol 2025i-D only. Table B–1. C-Link Protocol Commands Command Description...
Page 304
Reports the value for the actual flow (filter 1) in L/min flow mass Reports the value for the mass flow (filter 1) in L/min flow std Reports the value for the std flow (filter 1) in L/min Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 305
Reports/sets memory allocation for short records B-15 me (menu) Simulates pressing menu pushbutton B-18 menutext Displays the text of the menu item where the cursor is currently B-20 positioned Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 306
Reports/sets the timezone string for the network time protocol B-29 server Simulates pressing up pushbutton B-18 wind dir Reports the wind direction in degrees (0-360) wind speed Reports the wind speed in km/h Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Wait mode and is using the Basic sampling method. Send: flags Receive: flags 00120000 Figure B–1. Flags ambient pres This command reports the current value for the ambient pressure, in mmHg. Send: ambient pres Receive: ambient pres 761.8 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 308
This command reports the value for the filter2 pressure in mmHg. Send: filter2 pres Receive: filter2 pres 734.5 filter temp This command reports the value for the filter temperature in degrees C. Send: filter temp Receive: filter temp 18.8 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 309
2.40 wind dir This command reports the wind direction in degrees (0-360). Send: wind dir Receive: wind dir 269 wind speed This command reports the wind speed in km/h. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
24.1 14.9 4.9 3.2 -3.2 diag volt mib This command reports the diagnostic voltage measurements on the measurement interface board. The sequence of voltages is: Positive 24, B-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The following example sets the data treatment for concentrations in lrec as minimum. Send: set data treatment lrec min Receive: set data treatment lrec min ok Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-11...
Page 312
"flags" command. Send: erec Receive: erec 15:45 03-01-10 flags 130000 lrec srec lrec rec num srec rec num lrxy rec num srxy rec num B-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 313
These commands set the output format for lrecs and srecs, and erec data, according to Table B–2. The following example sets the lrec output format to ASCII with text. Send: set lrec format 1 Receive: set lrec format 1 ok Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-13...
= | 1 | 5 | 15 | 30 | 60 | These commands set the lrecs and srecs logging period to value in minutes. The following example sets the lrec logging period to 15 minutes. B-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 315
For more information on how to edit the scratch pad, see the “sp field” command. The following example copies the current list in scratch pad into the lrecs list. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-15...
Page 316
“set copy sp to lrec”, “set copy sp to srec”, or “set copy sp to stream” commands. Send: set sp field 1 5 Receive: set sp field 1 5 ok B-16 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Attaches time stamp to streaming data string frec This command retrieves the most recent filter record. If there is no data, the command returns “NO DATA”. Send: frec Receive: frec NO DATA Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-17...
Page 318
= | do | down | en | enter | he | help | le | left | me | menu | ri | right | ru | run | up | 1 | 2 | 3 | 4 | B-18 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
4-bit-per-pixel gray-scale image. Also note that BMP files are upside down relative to this data, i.e. the top display line is the last line in the BMP. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-19...
This command sets the screen’s level of contrast, according to Table B–4 The following example sets the contrast level to 50%. Send: set contrast 10 Receive: set contrast 10 ok Table B–4. Contrast Levels Contrast Level Level B-20 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 321
Send: set save params Receive: set save params ok time This command reports the current time (24-hour time). The following example reports that the internal time is 2:15:30 pm. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-21...
Note This command cannot be used when DHCP is on. Refer to the DHCP command that follows for additional information. ▲ Send: set addr gw 192.168.1.1 Receive: set addr gw 192.168.1.1 ok B-22 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 323
This command sets the network time protocol server address, where address consists of four numbers ranging from 0-255 inclusive, separated by “.”. Send: set addr ntp 10.209.43.237 Receive: set addr ntp 10.209.43.237 ok Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-23...
Page 324
Until you cycle the power, the address assigned by the DHCP server will still be used and reported as the current address. ▲ B-24 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
This command returns “bad cmd” if no host name has been set. Send: set host name analyzer01 Receive: set host name analyzer01 ok instr name This command reports the instrument name. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-25...
Page 326
“ok” as if the command has been accepted and acted upon, but it will not change the instrument lock status (this is for compatibility with systems expecting and “ok” response). B-26 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
The following example sets the instrument to power up in remote/locked mode. Send: set power up mode 1 Receive: set power up mode 1 ok Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-27...
This command responds with “feature not enabled” if the I/O expansion board is not detected. Send: set analog iout range 4 1 Receive: set analog iout range 4 1 ok Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-29...
2 3 ok Table B–10. Analog Voltage Output Range Values Output Range Range 0-1 V 0-100 mV 0-10 V 0-5 V 0 [cannot be set to this, but may report] Undefined B-30 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 331
The following example sets the digital output channel 4 to state 11. Send: set dout 4 11 open Receive: set dout 4 11 open ok Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-31...
The active state for digital outputs is open or closed. The active state for digital inputs is high or low. Channels that have not been assigned a variable index are not displayed. Send: list dout Receive: list dout B-32 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 334
These commands set the relay logic to normally open or closed for relay number value, where value is the relay between 1 and 16. The following example sets the relay no 1 logic to normally open. B-34 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
'\n'. Valid fields are: t - parse a time specifier (2 bytes) D - parse a date specifier (3 bytes) Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-35...
The datum identified by the value source can be printed as a string 's', hexadecimal 'x', decimal 'd', or floating point 'f', or binary 'b' number. Typically, bitfield extractions are only done for decimal or hexadecimal numbers. B-36 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
B- Indicates a button which pops up an input dialog prompting the user for a new value using the designated input format. The input format is specified from the 'B' through the subsequent semicolon. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-37...
NO using a d.ddd format." The string entered by the user is used to construct the output command. If the user enters, for example, '1.234', the constructed command will be 'set no coef 1.234'. B-38 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 339
The background value is taken from the 7th element, and the precision specifier is taken from the 8th. If the asterisk were not present, it would indicate instead that 8 digits after the decimal point should be displayed. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual B-39...
Sampler. This is achieved by using the following supported MODBUS commands. For details of the Partisol 2025i and 2025i-D MODBUS Protocol specification, see the following topics: “Serial Communication Parameters”...
0x00 to 0xFF (Passed back in response) A Slave address is not required in MODBUS over TCP/IP because the higher-level protocols include device addressing. The unit identifier is not used by the Sampler. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
If the returned output quantity is not a multiple of eight, the remaining bits in the final data byte will be padded Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 344
Here is an example of a request and response to read outputs 2–15: Request Field Name (Hex) Function 0x01 Starting Address Hi 0x00 Starting Address Lo 0x02 Quantity of Outputs Hi 0x00 Quantity of Outputs Lo 0x0D Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
0x0A, or binary 0000 1010. Output 15 is in the fifth bit position from the left, and output 11 is the LSB of this byte. The four remaining high order bits are zero filled. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 346
Here is an example of a request and response to read registers 10–13: Request Field Name (Hex) Function 0x03 Starting Address Hi 0x00 Starting Address Lo 0x09 No. of Registers Hi 0x00 No. of Registers Lo 0x04 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Note This function will not work if the Sampler is in Service Level mode. ▲ Request Function code 1 Byte 0x05 Starting Address 2 Bytes 0x0000 to maximum allowed by Sampler Output Value 2 Bytes 0x0000 or 0xFF00 Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Data Unit (PDU) addresses. Verify the coil number on your MODBUS master to ensure that it matches the coil number on the Sampler. ▲ Note Coil status 1 indicates active state. ▲ Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
FLOW VOLUME 40003&40004 FLOW2 VOLUME 2025i-D only 40005&40006 FLOW TOTAL 40007&40008 FLOW COVARIANCE 40009&40010 FLOW2 COVARIANCE 2025i-D only 40011&40012 VOLUME 40013&40014 VOLUME2 2025i-D only 40015&40016 AMBIENT PRESSURE 40017&40018 FILTER PRESSURE Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 350
ANALOG IN 5 (I/O Expansion Board installed) 40101&40102 ANALOG IN 6 (I/O Expansion Board installed) 40103&40104 ANALOG IN 7 (I/O Expansion Board installed) 40105&40106 ANALOG IN 8 (I/O Expansion Board installed) C-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Refer to the iPort Instruction Manual part number 102606-00 for detailed information. RPComm Because the use of AK Protocol has been enabled on the Partisol 2025i and 2025i-D Samplers, RPComm can be used for communications with the Samplers. Because of the updated operating platform compared with the original Partisol series Samplers, not all functionality is available when using RPComm.
3. Press the “Next >” button when the RPComm Welcome screen appears and follow the instructions to install the program. Note Refer to the RPComm manual for more instructions on installing and using RPComm. ▲ Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Partisol Sampler. Select the appropriate PRC list and save it to a file. Note Be sure to remember the folder the PRC list file is saved in. ▲ 4. Disconnect from the Thermo Fisher Scientific Website. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
4. On the RS232 Setup screen, ensure that the proper ser port settings are selected. These include the Baud rate, data bits, parity, stop bits and Rs-232/RS-485 selection. The default settings are 9600, 8, none, and Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
1. Initiate the RPComm software by selecting the “Start” button on your PC’s screen, highlighting “Programs,” and then highlighting “RPComm.” Choose the RPComm icon to begin executing the RPComm program (Figure D–3). Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Figure D–3. Executing RPComm 2. When RPComm begins executing, two screens are always displayed: the Communicator Main screen and the Connect List screen (Figure D–4). Figure D–4. Connect List Screen Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
(Figure D–5). Figure D–5. Connection Type Screen 2. From the Connection Type screen, select the “2000” button and choose “Settings” to display the AK Protocol Setup screen (Figure D– Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
RS232 port settings, these values must sometimes be changed. To check these values, from the Connect List screen, select the “Setup” pull down menu and choose “RS232.” Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
“Open Group” from the “File” pull down menu of the Communicator Main screen. A list of saved connection groups will be displayed in the Save Connection Group screen (Figure D–8). Choose the desired group and select “OK.” Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
If the connection is not successful, or if there is no instrument attached, then the serial number area will be blank or will display “99999”. ▲ Figure D–9. Main Screen D-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
However, if the user wishes to download data only from a particular date, for example, then the position of the storage pointer will need to be set. There are three storage pointers used in the Partisol 2025i and 2025i-D Samplers, one for each type of data (filter, interval and input).
2. From the RPComm Main screen (Figure D–9), select the Download Data tab. Refer to Figure D–12 Figure D–13 for a complete description of the control buttons displayed on the Download Data screen. D-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Communications Using iPort or RPComm Using RPComm Figure D–12. Download Data Screen Figure D–13. Control Buttons on the Download Data Screen Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-13...
Windows clipboard for use in other Clipboard applications. Use the following procedure to copy data to the Windows clipboard. 1. From the Partisol 2025i Main screen (Figure D–9), select the Filter Data, Interval Data or Input Data tab to display the desired data screen (Figure D–14,...
Communications Using iPort or RPComm Using RPComm Figure D–14. Downloaded Filter Data Figure D–15. Downloaded Interval Data Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-15...
Save Dialog box will appear again prompting the user for a file name for the next type of data. This will continue until all downloaded data have been saved. D-16 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
2. Check the box next to the day(s) of the week that you want the data to be downloaded. You may choose a single day each week or any other combination of days. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-17...
Scheduled Download box. During a scheduled download, RPComm assigns file numbers or names to the downloaded data according to the file numbers that already exist in the D-18 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
2. On the right portion of the Registers screen is a list of all the system registers. Using the scroll bar, examine the list of registers and place a checkmark next the registers that you wish to view. Or, if desired, select Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-19...
Modem communications can be used to remotely download data from the unit, check instrument operation and to change system variables. All iSeries instrumentation can be connected to a standard external modem for communications. D-20 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Also, some modems limit the length of the command string that may be sent. The command string can be broken into two or more segments and sent individually. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-21...
RPComm under the Windows 98 or Windows NT operating systems. ▲ 1. Set up the Sampler for direct communications. Refer to “Instrument Setup for Direct Communication” on page D-4. D-22 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Edit Selected Connection icon. The Connection Type screen will then be displayed (Figure D–21). Figure D–20. Connect List Screen Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-23...
56K Voice INT PnP.” The description will vary depending on factors such as your modem type, whether it is an internal or external modem and which COM port the modem is connected to. D-24 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
“Dialing String” box. This setting will be automatically configured by the system. 14. Select the “Dialing Setup >>” button to display the Dialing Properties screen (Figure D–24). Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-25...
“99999.” If the modem loses its connection, wait at least 1 minute before trying to establish another connection. This allows the modem to reset itself. ▲ D-26 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
1. Set up the unit for modem communications (“Standard Commercial External Modem Setup” on page D-21). Verify that modem communications to each instrument are successful before adding the serial switching device. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-27...
Page 380
8. The remaining portions of the screen should have been set while establishing the modem connection (step 1). Do not modify these settings. 9. Select “OK” to exit the AK Protocol Setup screen. This will display the Connection Type screen. D-28 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
(Figure D–20) and select the Connection icon on the tool bar. The connection to the instrument will be initiated and the proper command codes sent. When communication is established, the Partisol 2025i Main screen will be displayed (Figure D–25). Note If the connection is successful, the unit’s serial number will be displayed at the top of the screen.
Using RPComm Connecting a Series 1400a Use the following procedure to connect a Series 1400a monitor and a Partisol 2025i-D to a serial switching device. and a Partisol 2025i-D 1. Set up the 1400a monitor and Sampler for direct communications (“Instrument Setup for Direct...
11. Attach a phone line to the modem. 12. With the RPComm Connect List screen displayed (Figure D–20), select the name of one of the connections used to verify the direct Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-31...
Page 384
“Dialing String” box. This setting will be automatically configured by the system. 21. Select the “Dialing Setup >>” button to display the Dialing Properties screen (Figure D–24). D-32 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 385
D–20) and then select the Connection icon on the tool bar. The modem connection will now be initiated. When communication is established, the instrument’s main screen will be displayed. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-33...
Page 386
The Connection Type screen will then be displayed (Figure D–21). 35. From the Connection Type screen, ensure that the correct instrument type is selected and select “Settings” to display the AK Protocol Setup screen (Figure D–22). D-34 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 387
If the modem loses its connection, wait at least 1 minute before trying to establish another connection. This allows the modem to reset itself. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-35...
Using RPComm Connecting a Model 2000- Use the following procedure to connect a Partisol Model 2000-FRM Sampler and a Partisol 2025i/2025i-D Sampler to a serial switching device. FRM and a Partisol 2025i/2025i-D Sampler 1. Set up the Partisol 2000-FRM Sampler and Partisol 2025i/2025i-D Sampler for direct communications (“Instrument Setup for Direct...
Page 389
Connection Type screen will then be displayed (Figure D–21). 13. From the Connection Type screen, ensure that the correct instrument type is selected and select “Settings” to display the AK Protocol Setup screen (Figure D–22). Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-37...
Page 390
If the phone number is not correct, there is an error on this screen or you may have incorrectly entered the phone number on the AK Protocol Setup screen. D-38 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
25. From the Communicator Main screen (which is located just behind the Connect List screen), select the “Setup” pull down menu and choose “RS232.” The Communication Port Setup screen will appear. The settings on this screen should match the Partisol 2025i/2025i-D Sampler settings as shown in Figure D–27.
Page 392
Connection Command box. If a code is entered incorrectly, it can be deleted by selecting the code in the Connection Command box and pressing the Delete key on your computer’s keyboard. D-40 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 393
If the modem loses its connection, wait at least 1 minute before trying to establish another connection. This allows the modem to reset itself. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual D-41...
AK Protocol This appendix provides a description of the AK protocol commands that can be used to remotely control a Partisol 2025i/2025i-D using a host device such as a PC or a datalogger. AK protocol may be used over RS-232, RS-485, or Ethernet.
‘SE’ indicates that at least one of the command’s parameters is out of range or another error has occurred. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
AK storage pointer resides at the oldest record in the internal data logger. The current record type affected by the SSTO command is denoted by PRC register 31. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Up to 3 digits appended to the end of the response transmission according to <ETX> the entry for RS-Para 3. <ETX> <CR> <space> <LF> ASCII code 003. <ETX> <CR> <LF> Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 399
PRC is less than 3 characters long. <space> <CR> Space. <ETX> <LF> ASCII code 003. Current value of the variable referenced by the Ask Register command. NOTE: This value can be of varying length. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 400
Description No Err Error Description <ETX> ASCII code 003. <CR> Up to 3 digits appended to the end of <LF> the response transmission, according to the entry for RS-Para 3. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 401
Interval Data 2 in PRC 31 Input Data The instrument then transmits the number of storage records shown in response bytes 10 to 12 above. Each record is followed by <CR><LF>. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 402
RS-Para 3. variable referenced by Program Register Code in bytes 11 to 13 above. NOTE: The value entered may be of varying length. <ETX> ASCII code 003. Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 403
ASCII code 003. <ETX> ASCII code 003. <CR> Up to 3 digits appended to the end of the response transmission according to <LF> the entry for RS-Para 3. <ETX> <CR> <LF> Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual...
Page 404
LISTING OF FUNCTION CODES (xx) <LF> of the response transmission, according to the entry for RS-Para 3. <ETX> Stop Set Time <CR> Set Date Switch to "None" RS232 Mode <LF> E-10 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 405
250 to move forward by n records, and negative numbers <LF> such as -1000 to move backwards by n records. Do not right-fill. <ETX> ASCII code 003. Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-11...
Page 406
RS-Para 3. Up to 3 digits appended to the end <LF> of the response transmission, according to the entry for RS-Para 3. E-12 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Analog Output Current Channel 5 Code* Anytime Analog Output Current Channel 6 Code* Anytime Storage Download Type Code* 0 -2 Anytime Analog Input Reading 1 User defined User defined None Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-13...
Page 408
Digital Output PRC 9 Code* Anytime Digital Output PRC 10 Code* Anytime Digital Output Logic 1 Code* Anytime Digital Output Logic 2 Code* Anytime Digital Output Logic 3 Code* Anytime E-14 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 409
Average Filter Temperature 1 °C Never Average Filter Temperature 2 °C Never Current Wind Direction Never Average Wind Direction Never Current Wind Speed km/h 0 -180 Never Average Wind Speed km/h Never Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-15...
Page 410
Anytime Sample Filter Type A2 0 -26 Anytime Sample Filter Type A3 0 -26 Anytime Sample Filter Type A4 0 -26 Anytime Sample Filter Type A5 0 -26 Anytime E-16 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 411
Anytime Sample Filter Type B4 0 -9999999 Anytime Sample Filter Type B5 0 -9999999 Anytime Sample Filter Type B6 0 -9999999 Anytime Sample Filter Type B7 0 -9999999 Anytime Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-17...
Page 412
Anytime Sample Filter ID A6 0 -9999999 Anytime Sample Filter ID A7 0 -9999999 Anytime Sample Filter ID A8 0 -9999999 Anytime Sample Filter ID A9 0 -9999999 Anytime E-18 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 413
Anytime Sample Filter ID B8 0 -9999999 Anytime Sample Filter ID B9 0 -9999999 Anytime Sample Filter ID B10 0 -9999999 Anytime Sample Filter ID B11 0 -9999999 Anytime Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-19...
Page 414
Sample Filter Cassette ID A10 0 -9999999 Anytime Sample Filter Cassette ID A11 0 -9999999 Anytime Sample Filter Cassette ID A12 0 -9999999 Anytime Sample Filter Cassette ID A13 0 -9999999 Anytime E-20 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 415
Sample Filter Cassette ID B12 0 -9999999 Anytime Sample Filter Cassette ID B13 0 -9999999 Anytime Sample Filter Cassette ID B14 0 -9999999 Anytime Sample Filter Cassette ID B15 0 -9999999 Anytime Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-21...
Note A status code value showing “00000001” indicates that there are no status codes (NONE). ▲ PRC 5: Status Code (H)1 NONE (H)2 MEMORY (H)10 POWER FAILURE (H)20 FLOW (H)40 FLOW2 E-24 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 420
ANALOG IN 8 PRC 31: Storage Download Type FILTER DATA INTERVAL DATA INPUT DATA PRC 60 – PRC 65: Analog Voltage Output Range 1 volt 100 millivolt 10 volt 5 volt E-26 Partisol 2025i and Partisol 2025i-D Instruction Manual Thermo Fisher Scientific...
Page 421
SERVICE EXTERNAL ALARM 1 EXTERNAL ALARM 2 EXTERNAL ALARM 3 PRC 82 – PRC 91: Digital Output Active State NORMALLY CLOSED NORMALLY OPEN PRC 162: Default Filter Type Space Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-27...
Page 423
DIG IN8 PRC 434: Sampling Action Under Serial Control Stop Sampling (Sampling Not Active) Exchange Cassette, Goes to State 2 (Sampling Active) Activate Sampling (Sampling Active) PRC 437: Auto Run Thermo Fisher Scientific Partisol 2025i and Partisol 2025i-D Instruction Manual E-29...
Need help?
Do you have a question about the Partisol 2025i and is the answer not in the manual?
Questions and answers