Water Connections; Feed Water Connections - Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Advertisement

Water Connections

Feed Water Connections

Thermo Scientific
Accudispense
control cable
Pump interlock
Remote 1
Remote 2
Feed water inlet
Atmospheric drain
Figure 5-6. Left Side View
Caution Do not connect feed water until directed to do so during "Initial
Cleaning."
s
1. Locate the length of 3/8" O.D. tubing provided with the unit with a
quick disconnect insert on one end and a 3/8" (.95 cm) O.D. X 1/4"
(.64 cm) NPT tubing adapter on the other.
2. Install the tubing adapter onto your incoming water line. We
recommend a customer supplied shut off valve be installed in your feed
water line. Do not connect the feed water to your Nanopure. You will
connect the feedwater during the "Initial Cleaning" procedure
explained later in this manual.
Note The cartridge pack and ultrafilter must be installed before connecting
the water line.
s
3. If a pressure switch is to be used, see the "Installing a Float or Pressure
Switch" section for more information.
Overflow float
connection
Figure 5-7. Right Side View
Nanopure Life Science (UV/UF)
Section 5
Installation
5-7