Problem
Water is leaking between the manifold and
the cartridge pack.
Water is leaking at atmospheric drain connec-
tion
Thermo Scientific
Possible Causes
Wing head screws are not tight enough.
New cartridge pack: o-ring missing from post
(each of the 4 posts must contain an o-ring.)
Debris on o-ring or bottom of unit manifold.
Tubing not fully inserted.
Solutions
Tighten wing head screws on unit manifold.
Disconnect inlet water and depressurize sys-
tem. Next, disconnect the manifold and
verify/replace missing o-ring. Reinstall car-
tridge pack as indicated in "Cartridge Pack
Replacement."
Inspect o-rings and bottom of unit manifold.
Remove any objects with a lint-free cloth.
Insert tubing into connection until it bottoms
out.
Nanopure Life Science (UV/UF)
Section 13
Troubleshooting
13-3